XI. Error Codes

1.  OUTZ

 - Zero range has shifted. Perform calibration 

adjustment. If the problem persists, the load cell may 

be permanently damaged due to overloading.

2.  O-LD

 - Maximum capacity exceeded. Remove the 

extra weight to avoid damaging the load cell.

3.  LO 

- Batteries are low. Replace the batteries.

4.  UNSE 

- The scale is unstable. Use the scale in a 

more stable position.

ACP-200

ACP-500

Max. Capacity

200g

500g

Readability

0.01g

0.1g

Power

2 x AAA Batteries

Dimensions

2.8 x 4.8 x 0.8 in.

Platform Dimensions

2.5 x 3.0 in.

Warranty

10 Year Limited Warranty