XI. Error Codes
1. OUTZ
- Zero range has shifted. Perform calibration
adjustment. If the problem persists, the load cell may
be permanently damaged due to overloading.
2. O-LD
- Maximum capacity exceeded. Remove the
extra weight to avoid damaging the load cell.
3. LO
- Batteries are low. Replace the batteries.
4. UNSE
- The scale is unstable. Use the scale in a
more stable position.
ACP-200
ACP-500
Max. Capacity
200g
500g
Readability
0.01g
0.1g
Power
2 x AAA Batteries
Dimensions
2.8 x 4.8 x 0.8 in.
Platform Dimensions
2.5 x 3.0 in.
Warranty
10 Year Limited Warranty