be weighed. The display will show the net weight of
the contents.
4. To return to normal weight display (gross weight),
remove any items from the weighing platform and
press the
TARE
key again to re-zero the scale.
VII. Calibration
Calibration and adjustment are performed at the factory.
However, the weighing range can shift slightly during
shipping. Also, due to changes in local acceleration of gravity
around the world, adjustment is recommended when moving
the scale from one location to another. To check calibration:
1. Place the required calibration weight (see chart) on
the scale and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart below), adjustment is required.
Model
Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
5K-BOWL
2g
5kg
OIML M2
VIII. Adjustment
1. Turn the scale ON.
2. Tap the
ON/OFF
key four times and hold it on the
fourth tap.
3. The display will show a random number (ISN). Make
sure no weight is on the scale and press the
ON/
OFF
key again.
4.
The display will flash “Zero” followed by the required
calibration weight (5000g/5kg).
5. Place the required weight on the center of the
weighing platform.