23
If a value has a decimal point whose position can be changed, press the
SELECT button until the underscore is under the decimal point. Press the
DOWN button to move the decimal point to the left and press the UP button to
move the decimal point to the right.
NOTE: Not all programming screens allow the decimal point location to be changed such as any
milliamp value.
Once a variable has been changed to the desired value, press the ENTER button
to accept the value and move to the next screen or programming value. If an
incorrect value is programmed, a warning or error screen will appear. Press any
button to exit a warning screen, but you must correct an error before moving on.
RATE, TOTAL and GRAND TOTAL Scaling
Figure 11: Programming Mode
K-factor (scaling factor)
The MX 9000 uses one K-factor for each channel for scaling displays. This
value is always entered in units of Pulses per Gallon. Once the rate and total
units are selected, the MX 9000 uses internal calculations to automatically
correct the displayed values to match the user selected units. When the correct
K-factor value has been programmed, press ENTER to continue.
RATE unit
Use the UP and DOWN buttons to scroll through the available units to scale the
RATE screen.