ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
±
8kV contact
±
2 kV,
±
4kV
±
,
±
8 kV,
±
15 kV air
Radiated RF EM
fields
IEC 61000-4-3
a)
10 V/m b) 80MHz - 2.7 GHz 80%
AM at 1kHz
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
COMPLIANT
NOTE: Further information about distances to be maintained
between portable and mobile RF communications equipment
(transmitters) and the BPM29 can be requested from supplier using
the contact information provided in this manual. However, it is
advisable to keep the electromechanical aerosol equipment at an
adequate distance of, at least, 0.5 m from mobile phones or other
RF communications transmitters to minimise possible interference.
RATED power
frequency magnetic
fields.
IEC 61000-4-8
30 A/m c)
50 Hz or 60 Hz
a) The equipment is suitable for use in Home Health Environments and Professional Health Care
Environments limited to patient rooms and respiratory treatment facilities in hospital or clinics. The more
restrictive IMMUNITY acceptance limits have been considered and applied.
b) Before modulation is applied.
c) This test level assumes a minimum distance of at least 15 cm between the ME EQUIPMENT or ME
SYSTEM and sources of power frequency magnetic fields.
EMC Tables