Installation and Setup
Mounting the System
The Pico unit can be mounted to a wall or placed on a sturdy bench. Avidity Science recommends wall-
mounting the unit to be sure that a connection to the lower dispenser-tap port is accessible. Refer to
on page 5 for information on placement of the Dispenser Tap.
Bench Placement
The bench must be able to hold 60 kg [132 pounds] and located on an flat, level surface. The surface of the
bench must be resistant to applied chemicals such as disinfectants, acids, dyes, and solvents.
Wall Mounting
The wall-mounting kit is separate. Use the included instructions as a supplement to this procedure to
mount the Pico unit. In addition to the kit, these are the tools required to mount the Pico Water Purifica-
tion unit to a wall.
•
Wall Mounting kit, EI #1330-3000-001 (PICOWB)
•
Pozi screwdriver
•
Pen/pencil for marking wall
•
Level
•
Drill with 10 mm drill bit
•
4 mm Allen Head wrench/Hex key
IMPORTANT:
It is recommended that at least two people assist in mounting the unit.
Follow this procedure. Refer to Figure 3.
1.
Measure the width of the Pico enclosure. This measurement (
A
) will be used as the spacing
between mounting brackets.
2.
Position and level the left mounting bracket (
L
) vertically at the location desired mounting loca-
tion. Mark the bracket location and the four mounting hole locations on the wall.
3.
Using a drill with a 10 mm masonry bit, drill four mounting holes at the marked locations deep
enough to insert wall anchors.
4.
Insert the wall anchors.
5.
Position the left mounting bracket (
L
) holes over the anchors and secure the bracket with four
anchor screws.
6.
Position the right mounting bracket (
R
) from the left bracket the width measurement (
A
) acquired
7.
Make sure the right mounting bracket (
R
) is vertically level and level with the left bracket. Mark the
bracket location and the four mounting holes on the wall.
CAUTION!
Personal injury. The reservoir must be empty when mounting the
Pico unit. A reservoir with water adds extra weight that can result in personal
injury when lifting the unit onto the mounting brackets.
Summary of Contents for Pico PICO10T2
Page 1: ...Pico Water Purification System Operation Manual OP 000100 Revision D August 2018...
Page 22: ...Pico Water Purification System Operation Manual Avidity Science LLC 16 Installation and Setup...
Page 34: ...Pico Water Purification System Operation Manual Avidity Science LLC 28 Maintenance...
Page 38: ......
Page 40: ...Pico Water Purification System Operation Manual Avidity Science LLC 34 Replacement Parts...