14
3700LP Bench Scale Series User’s Manual
DISP
Shows version and revision of program, followed by a segment
test and countdown.
RA
Performs a nondestructive test of all locations of RAM in the
processor and displays
PASS
or
FAIL
.
RO
Performs a checksum of all ROM locations in the processor.
This value is compared with a value stored in ROM and displays
PASS
or
FAIL
.
I/O
Install a loopback connector with transmit and receive jumped
together on the RS-232 output connector. The display will show
PASS
or
FAIL
.
HRESA
Displays ten times normal resolution with AZT enabled and the
most significant digit of weight not shown. Press
TEST
to stop
this test.
HRESN
Displays ten times normal resolution with AZT disabled and
most significant digit of weight not shown. Press
TEST
to stop
this test.
CAP
Displays the calibrated capacity and division size of the scale.
UNITS
Displays the current unit of measure selection(s).
LOC-G
Shows the local gravity compensation value.
CAL-G
Shows the gravity compensation value for the Fairmont, MN
manufacturing site.
The following are viewable configuration settings. None can be
changed:
FILTR
Filter setting - FAST, SLOW
BACLT
Backlight setting - ON, OFF, AUTO
BAUD
Baud rate setting - 1200 to 19.2K
PROT
RS-232 protocol setting
AUTOP
Autoprint setting - OFF, AUTO1, AUTO2
E-INP
External input setting - ZERO or PRINT
SLEEP
Sleep setting - NONE, 5, 30, 60 minutes
P-OFF
Power off setting - ON, OFF
SEAL
Seal setting - SEALED, UNSEALED