6.

 

Specifications 

 

 

AvaLight-
LED355/380 

AvaLight-
LED400/410/430 

AvaLight-
LED450/470/490 

AvaLight-
LED530/590/780 

Spectral Range* 

 

355/380 nm 

400/410/430 

nm 

450/470/490 

nm 

530/590/780 

nm 

FWHM (nm) 

15 nm 

11 nm 

30 nm 

30 nm 

Optical power 600 µm 
fiber 

 

10 µWatt 

25 µWatt 

25 µWatt 

25 µWatt 

Connector 

 

SMA 905 

Power supply 

12 VDC, 40 mA 

Dimensions, weight

 

175 x 110 x 44 mm, 480 grams 

* Other wavelengths available on request 

 

6.1

 

Ordering Information 

 

AvaLight-LED-XXX

 

Light Emitting Diode Light source, specify wavelength XXX 

AvaLight-LED-XXX-RM

  Rackmount version of the Light Emitting Diode Light source, specify wavelength 

XXX 

AvaLight-LED-CON 

LED light source control unit with electrical connector to LED, needs extra PS-
12V/1.0A and interface cable. 

CUV-LED-XXX 

LED holder for Cuvette, specify LED wavelength XXX 

CUV-DA

 

Direct-attach cuvette holder for AvLight-DHc/XE/LED 

IC-DB26-2 

Interface cable AvaSpec-USB2 platform to AvaLight-LED for PWM  

PS-12V/1.0A

 

Power supply 100-240 VAC/12VDC, 1.0 A for AvaLight-LED 

 
 
 

Summary of Contents for AvaLight-LED

Page 1: ...NEED TECHNICAL SUPPORT Scan the QR code or visit www avantes com support We are happy to help you AVALIGHT LED LIGHT SOURCES Operation and Installation Manual...

Page 2: ...oduct specifications and descriptions in this manual are subject to change without notice Avantes BV will not be responsible for errors and omissions in this manual or for direct or incidental damages...

Page 3: ...start 4 2 2 12VDC Power plug 5 2 3 Settings for the Pulsed Mode 5 3 AvaLight LED CON 6 3 1 Quick start 6 3 2 12VDC Power plug 7 3 3 Settings for the Pulsed Mode 7 4 Changing the LED of the AvaLight LE...

Page 4: ...light to the near infrared it depends on the choice of the wavelength for the LED The AvaLight LED CON is a LED control unit with a connector to a LED The AvaLight LED CON works in combination with th...

Page 5: ...rce when the power is turned on OFF TTL ON Switch Choose operating mode with the mode switch Switch position PWM signal Light source LED OFF X OFF OFF TTL Not connected ON OFF High ON ON Low ON OFF ON...

Page 6: ...e electrical characteristics for the input signals Electrical characteristics Name Low level detection High level detection Maximum voltage PWM 0 7V 0 7V 5 0V 2 2 12VDC Power plug The power supply spe...

Page 7: ...ng mode with the mode switch Switch position PWM signal Light source LED OFF X OFF OFF TTL Not connected ON OFF High ON ON Low ON OFF ON X ON ON LED connector Connect the CUV LED XXX to this connector...

Page 8: ...e electrical characteristics for the input signals Electrical characteristics Name Low level detection High level detection Maximum voltage PWM 0 7V 0 7V 5 0V 3 2 12VDC Power plug The power supply spe...

Page 9: ...LED holder and LED connector 7 Slightly bend the electrical wires out of the LED connector 8 Take out the LED from the LED holder 9 Replace the LED make sure the Anode and Cathode are connected in the...

Page 10: ...equests concerning your AvaLight LED CON or the AvaSoft software please go to http www avantes com In case you have any questions or need support please send us an email via support avantes com 5 2 Av...

Page 11: ...ams Other wavelengths available on request 6 1 Ordering Information AvaLight LED XXX Light Emitting Diode Light source specify wavelength XXX AvaLight LED XXX RM Rackmount version of the Light Emittin...

Page 12: ...ion 1 00 11 Avantes BV www avantes com info avantes com 7 Appendix A Start AvaSoft and press the settings button see below The menu as shown below wills pop up Press the light source control tab First...

Page 13: ...ID 020405 Revision 1 00 12 Avantes BV www avantes com info avantes com Please execute the following steps 1 2 and 3 In the pull down menu you can choose what mode is required Than press ok 1 2 3...

Reviews: