Autonics TPS20 Series Manual Download Page 1

TPS20 

Series

F-7

 

DC4-20mA analog signal (2-wire) transmission by 
measuring pressure of liquid, gas, and oil.

 

High accuracy (±0.3

% 

F.S.) with stainless steel 

diaphragm for various measurement

 

Various model for installation environments

  : Head type, DIN connector type,  

  connector cable type

 

Built-in zero-point, span adjustment (head type)

Features

Please read “Safety Considerations” in operation manual 

before using this unit. 

Ordering Information

[Head type] [DIN connector type]

[Connector cable type]

G

1

5

F8

TPS20

② ③ ④ ⑤

(0 to 5kgf/cm

2

)

 

1: Write the desired pressure range and it is the default of user pressure range. (

select "Z" at ④Pressure range)

 

For ordering cable, order as CID3-2, CID3-5, CLD3-2, CLD3-5. (sold separately)

Description

① 

Item

TPS20

Pressure Transmitter

② 

Measurement presssure

G

Gauge pressure

A

Absolute pressure

③ 

Cable

1

Head type

2

DIN connector type

3

Connector cable type

④ 

Pressure range

Gauge pressure

Absolute pressure

1

0 to 0.2kgf/cm

-

2

0 to 0.5kgf/cm

2

-

3

0 to 1kgf/cm

2

 

0 to 1kgf/cm

2

 

4

0 to 2kgf/cm

2

 

0 to 2kgf/cm

2

 

5

0 to 7kgf/cm

2

 

0 to 7kgf/cm

2

6

0 to 10kgf/cm

2

0 to 10kgf/cm

2

 

7

0 to 20kgf/cm

2

 

0 to 20kgf/cm

2

8

0 to 35kgf/cm

2

 

0 to 35kgf/cm

2

 

9

0 to 70kgf/cm

2

 

-

A

0 to 100kgf/cm

2

-

C

0 to 200kgf/cm

2

 

-

F

0 to 300kgf/cm

2

 

-

H

0 to 350kgf/cm

2

 

-

M

-760mmHg to 0kgf/cm

2

 

-

O

-760mmHg to 1kgf/cm

2

 

-

Q

-760mmHg to 7kgf/cm

2

 

-

V

-760mmHg to 10kgf/cm

2

 

-

X

-760mmHg to 20kgf/cm

2

 

-

Y

-760mmHg to 35kgf/cm

2

-

Z

Others

⑤ 

Pressure port

P2

R1/2 (with adapter, PT)

P8

R3/8 (with adapter, PT)

F8

G3/8 (standard, PF)

ZZ

Others

⑥ User pressure range

User pressure range

1

Summary of Contents for TPS20 Series

Page 1: ...ID3 2 CID3 5 CLD3 2 CLD3 5 sold separately Description Item TPS20 Pressure Transmitter Measurement presssure G Gauge pressure A Absolute pressure Cable 1 Head type 2 DIN connector type 3 Connector cable type Pressure range Gauge pressure Absolute pressure 1 0 to 0 2kgf cm2 2 0 to 0 5kgf cm2 3 0 to 1kgf cm2 0 to 1kgf cm2 4 0 to 2kgf cm2 0 to 2kgf cm2 5 0 to 7kgf cm2 0 to 7kgf cm2 6 0 to 10kgf cm2 0...

Page 2: ...Connector cable type The standard pressure port for above is G3 8 Dimensions Connections Connection Cable Sold Separately M12 Ø14 8 L m Ø5 40 M12 L m Ø5 27 5 Ø14 8 32 CLD3 2 CLD3 5 CID3 2 CID3 5 unit mm Model L m Meterial CID3 2 2 PVC CID3 5 5 Model L m Meterial CLD3 2 2 PVC CLD3 5 5 21 14 34 71 106 G3 8 84 24 21 31 28 42 104 8 G3 8 46 5 2 7 21 31 62 10 8 M12 Connector G3 8 22 15 35VDC brown V blu...

Page 3: ...inute Vibration 1 5mm amplitude at frequency of 10 to 55Hz for 1 min in each X Y Z direction for 2 hours Shock 95m s2 Tightening torque Industrial plug over 5N Pressure port G3 8t standard R3 8 R1 2 Environment Ambient temp 10 to 70 storage 10 to 70 Ambient humi 5 to 95 RH storage 5 to 95 RH Materials Sealing diaphragm connection stainless steel type 316 O ring fluoro rubber Connection Case struct...

Page 4: ...essure slot cleanliness inside and corrosion state Short each terminal and check the insulation resistance between the case and power at 100VDC over 10MΩ Check zero span adjustment and linearity by pressure standards When removing a sensor for maintenance follow the below instructions Replace an O ring which is used once Be sure that diaphragm part is not damaged In case of head type for connectin...

Reviews: