      det$ = "Probe AD-17"
      unit$ = "cps"
    CASE 18
      det$ = "Probe AD-18"
    CASE 19
      det$ = "Probe AD-19"
      unit$ = "cps"
    CASE 20
      det$ = "internal GM"
    CASE 21
      det$ = "AD-t low DR"
    CASE 22
      det$ = "AD-t high DR"
  END SELECT
RETURN
' ---------------------------------------------------------------------
'       Handler for transmission errors
' ---------------------------------------------------------------------
RecvErr:
IF ERR = DEVERR THEN PRINT TIME$; " Transmission Error": RESUME MainLoop
ON ERROR GOTO 0

2.2 »Control«

This output switches between the two GM counters in the telescopic probe AD-t, where Control = high

(4.75V) selects the low dose rate GM counter, and Control = low (GND) selects the high dose rate GM

counter. If any other or no probe is connected, Control will always be high.

Certain software versions of the 6150AD (e.g., the Swedish version V-36.04) make additional use of the

Control output: With the AD-t probe, Control will operate as described above. With any other or no probe

the Control output will signal dose rate alarm: Control = high indicates that dose rate alarm is off, and

Control = low indicates that dose rate alarm is on. This allows triggering of external alarming devices.

However, one has to consider that the Control output can only drive very small loads due to its internal

structure:

The microprocessor port can sink only a few mA and can source almost no current, because the high state

is achieved through a weak pull-up resistor of 10-100 kOhms depending on the manufacturer of the

microprocessor. The Control output can therefore only drive high impedance low capacity inputs like

CMOS gates as shown in the drawing above. Such buffering is strongly recommended before connecting

some external device.

micro-

processor

port

1 kOhm

6150AD

4.75V
Control

GND

buffered

output

(e.g. 4049, 4069, 40106)

(inverted)

CMOS gate 4000 series

06/2003

Technical Manual 6150AD Probe Connector

Page 6 of 14

Daimlerstrasse 27

Telefon  +49(0)6203-9503-00 http://www.automess.de

USt.-ID-Nr. / VAT ID No.

D-68526 Ladenburg Telefax   +49(0)6203-9503-29 eMail: [email protected] DE 144 458 420

Summary of Contents for 6150AD

Page 1: ...Data Inputs 8 3 1 PL and S E Lines 8 3 2 Ext Input 10 3 3 Probe Characteristics 11 Daimlerstrasse 27 Telefon 49 0 6203 9503 00 http www automess de USt ID Nr VAT ID No D 68526 Ladenburg Telefax 49 0...

Page 2: ...low an will be discussed in more detail later GND Ground potential Note that there is no pin for GND The cable shield serves both as shield and as ground connection The aluminium housing of the 6150AD...

Page 3: ...ed Ohms Its driving capability is therefore at least equivalent to RS232 interfaces Nevertheless cables should be restricted to maximum lengths as with RS232 interfaces 15m at 9600Bd 30m at 4800Bd etc...

Page 4: ...xponent 6 block check character The second byte type of detector 6150AD contains Bit 0 5 detector in use 0 probe 6150AD 0 7 probe 6150AD b 15 probe 6150AD 15 17 probe 6150AD 17 18 probe 6150AD 18 19 p...

Page 5: ...e of detector 6150AD bc sonde mantlo ASC INPUT 1 1 low order mantissa bc bc XOR mantlo manthi ASC INPUT 1 1 high order mantissa bc bc XOR manthi expon ASC INPUT 1 1 exponent bc bc XOR expon IF expon 1...

Page 6: ...igh indicates that dose rate alarm is off and Control low indicates that dose rate alarm is on This allows triggering of external alarming devices However one has to consider that the Control output c...

Page 7: ...put battery voltage of 7 0V no load RL 82 5 Ohm Uin V Uout V Iin mA Uout V Iout mA efficiency 5 0 4 782 52 0 4 320 52 4 87 5 5 4 785 55 0 4 651 56 4 87 6 0 4 789 52 5 4 683 56 8 84 7 0 4 795 46 5 4 72...

Reviews: