Automation Dr. Nix QNix 1500 Instruction Manual Download Page 3

3900 values can be stored and statistically evaluated. 
Optionally, a special software for statistical evaluation can be 
supplied. 
The 

QNix

®

 1500 

complies with national (DIN) and international 

(ISO, BS, ASTM) standards and regulations: DIN 50981, 
50984; ISO 2178, 2360, 2808; BS 5411(3, 11), 3900 (c, 5); 
ASTM B499, D1400, D7091 

Maintenance and Handling 

This gauge incorporates the latest electronic technology in 
order to fulfill different measuring tasks. This high precision 
instrument covers a wide range of applications, yet it is robust 
and easy-to use in every-day practice. If operated and treated 
with care, it will serve you many years. 
Do not let it drop and protect it from dirt, dust and other 
substances. After use, the gauge should be stored in its 
protective carrying case. 
Substantial changes in temperature can influence the 
readings, as it is the case with all precision gauges. Do not 
expose the gauge to strong, direct sunlight or to other sources 
of energy causing temperature shocks. 
Protect the gauge from moisture, chemically aggressive 
substances and/or gases. 
The casing is resistant against most solvents. However, a 
resistancy against all chemical substances cannot be 
guaranteed. Use only a damp, soft cloth to clean the casing. 
Accurate readings can only be obtained, if the measuring 
probes are clean. Check the probes regularly and remove 
existing residues such as lacquer etc. from the ruby probe tip. 
If the gauge will not be used for a long period, remove the 
battery in order to prevent leakage and therefore destruction 
of the gauge. 
In case of a malfunction, please do not try to repair the gauge 
yourself. Our repair service will be glad to help at any time. 

Measuring Range 

0 – 5000 

μ

m or 0 – 200 mil in one measuring range 

Gauges without memory can be switched from 

μ

m to mil by 

operating the two switches on the sides of the instrument. 

3

Summary of Contents for QNix 1500

Page 1: ...1 Instruction Manual Coating Thickness Gauge 1500 Optionally with Memory...

Page 2: ...s this gauge offers Would you like information on the other gauges in our coating thickness measurement line or do you have any questions concerning coating thickness measurement Please do not hesitat...

Page 3: ...adings as it is the case with all precision gauges Do not expose the gauge to strong direct sunlight or to other sources of energy causing temperature shocks Protect the gauge from moisture chemically...

Page 4: ...ttery The battery is located at the wide end of the gauge If the blinking sign BAT appears on the display it is advisable to change the battery but it is still possible to measure for some time Memori...

Page 5: ...on the same spot the measured value may not always be 0 00 due to surface roughness dirt scratches etc The gauge is delivered with two protective rings around the probe for better positioning and prot...

Page 6: ...th a built in memory and an interface RS 232C offering the following features a Memory for up to 3900 values b Block segmentation up to 999 blocks Block No 1 Indication b001 Block No 2 Indication b002...

Page 7: ...the Memory Mode It is possible to store the measuring values in different blocks in order to distinguish the different tasks To set up a new block e g Block 2 push both slides simultaneously forward...

Page 8: ...o get any statistical value The display will then read AE Then the gauge automatically returns to the memory mode Deleting Measured Values While operating in the Memory Mode the QNix 1500 offers the p...

Page 9: ...change Information the measured values will remain stored 6 bloc Displayed alternating with the block number 7 End Memory or block capacity out of space 8 N a alternating with Fe or NFe number of meas...

Page 10: ...10 0 99 9 mil 1 mil in the range of 100 200 mil Repetitive accuracy 1 m 2 of reading in the range of 0 999 m 3 5 of reading in the range of 1000 5000 m Minimum object size 10 x 10 mm2 or 0 4 x 0 4 Mi...

Page 11: ...11...

Page 12: ...03 09 12 Automation Dr Nix GmbH Co KG Robert Perthel Str 2 50739 K ln Germany Tel 49 0 221 91 74 55 0 Fax 49 0 221 17 12 21 info automation de www automation de...

Reviews: