TROUBLESHOOTING
Please refer to the table below to troubleshoot any
problems.
Low atomization
.
• Check to ensure there is water in tank.
• Make sure water tank is firmly in place.
• Clean mesh as recommended in the instructions.
• Clean electrodes on control unit and water tank.
Power light goes out immediately after powering on.
• Make sure water cup is firmly in place.
• Clean electrodes on control unit and water cup.
Power indicator light does not come on.
• Ensure that the AC adapter is properly connected with
the main device.
• Check to see if battery pack is properly inserted in
control unit.
Power lights on but no atomization.
• Clean mesh as recommended in the instructions.
• Check mesh to see that it is not damaged.
• Clean electrodes on control unit and water tank.
• Check fluid level.
• Re-install the battery pack.
• Tilt unit forward.
Steam Inhaler shuts off during use.
• Make sure water cup is firmly in place.
• Clean electrodes on control unit and water cup.
• Re-install the battery pack.
• The internal temperature is overheating. Turn off the
device and cool it down for 10-20 minutes until the device
is completely cool.
• The treatment period has exceeded 25 minutes. To
activate the device, power the ON/OFF button for
another treatment.