with the “+” side
up, matching to
the “+” on the
battery compartment door.
Step 3: Swing the door
into the closed position.
The compartment should
close easily. Do not force it.
If resistance is noted, check
that the battery is inserted
correctly.
The life of your hearing
instruments' batteries will vary
depending on length of daily
use, listening environments, and how your
hearing instruments are programmed. When
the hearing instrument batteries go dead,
remove them immediately and replace them
with new batteries. Batteries may be purchased
through your hearing healthcare professional.
When your hearing instruments are not in use,
open the battery compartment to prevent
excessive battery drain and ventilate the
instruments.
Important: Always discard used batteries. Small
batteries can be harmful if swallowed. Keep
batteries out of the reach of pets and small
children. In case of ingestion, contact your
physician or call the National Button Battery
Hotline collect at (202) 625-3333.
6
b a t t e r y
i n f o r m a t i o n
The low battery detection system continuously
monitors battery life.
When the battery becomes low, the low
battery warning sequence consisting of 3
double beeps is initiated. After the warning
beeps are complete, normal operation resumes.
The low battery warning sequence will repeat
every 10 minutes until the battery becomes
critically low. At that time, the end of the
battery life warning sequence consisting of 6
double beeps is issued. The device is then
muted and put into low power sleep mode.
Battery replacement will then be required.
The volume of the low battery indicator can be
adjusted by your hearing healthcare provider.
To replace your hearing instrument battery,
follow these steps:
Step 1:
Lift the notch
at the edge of the
compartment and
swing the door open.
Remove the battery.
Step 2:
Remove the “brown”
colored tab on the 312
battery, slide in the battery
5
flx manual 3/6/09 3:56 PM Page 11