ATX Q series Installation & Operation Manual Download Page 9

          Q-Series

®

 Amplifiers – QPAIR Redundant Amplifier System – Installation & Operation Manual 

2-1

CHAPTER 2: DESCRIPTION OF OPERATION

DESCRIPTION OF OPERATION

2.  Description of Operation

(Refer to Figure 2 and Figure 6 for location of switches, push buttons, and LEDs)

RF output levels of both the primary and the secondary amplifiers are sampled and this sample is converted to a voltage 
value. At initial power on or when the reset button is pushed, the system is switched to run on the primary amp (AMP 1). The 
yellow LED will flash slowly indicating the reference levels need to be set. By pressing the SET REF button, the microcontroller 
samples the output values of both the primary amp and the secondary amp and stores the values as references. These 
values will be used to compare with the sampled output values of both amplifiers to confirm proper levels. At any time during 
the operation, one can always press RESET to reset the microcontroller program. Then push SET REF to store the new 
sampled values.

During normal operation, if both outputs are within 3 dB (factory programmable) of the reference values, no action will be 
taken. If the primary amp is down 3 dB or more and the secondary amp is not, the system considers the primary amp failed 
and will switch to the secondary amp, set the alarm, and set the red failure LED on. The microcontroller cannot perform any  
additional switches because the detectors are unable to sample the signal when the secondary amplifier is activated (see 
Figure 6). You will need to troubleshoot or replace the primary amplifier. See Appendix 3 for troubleshooting instructions 
or Appendix 2 for replacing the amplifier module. After the amplifier problem is resolved, you need to reset the system by 
pushing the RESET button. Set the levels according to the setup procedure and store levels by pushing the SET REF button.

If both outputs are down 3 dB or more, the system considers the input is low and will set the red LED flashing quickly and 
set the alarm. After resolving the low input level problem, set the system output levels according to the setup procedure. To 
restore normal operation push the RESET button and then push the SET REF button.
If the system is running on the primary amp and the secondary amp level is down 3 dB or more, the system considers the 

secondary amp failed and will set the alarm and set the red LED flashing slowly. You need to troubleshoot or replace the 

secondary amplifier. See Appendix 3 for troubleshooting instructions or Appendix 2 for replacing the amplifier module. After 

the amplifier problem is resolved, you need to reset the system by pushing the RESET button. Set the levels according to the 

setup procedure and store levels by pushing the SET REF button.
The signal level within the bandwidth of the primary amplifier coming into the RF switch compartment should be limited to 

not more than 50 dBmV per channel. However, if only part of the TV channel spectrum is used, the maximum signal level per 

channel may be higher than 50 dBmV and the minimum signal level should be higher than 20 dBmV. If needed, contact ATX 

Networks for assistance in calculating the correct levels.
The recommended primary or secondary amplifier output level is between 20 dBmV and 50 dBmV with all the TV channels 

loaded within the passband. The secondary amplifier operates 25 dB lower than the primary amplifier during normal operation 

and is switched to full output level when a switch to the secondary amplifier is made. The reduced level of the secondary 

amplifier during normal operation adds to the isolation between the outputs of the two amplifiers.
The system stores the reference levels to the non-volatile EEPROM memory in the microcontroller. When a power outage 

occurs and power is restored, the QPAIR will wait for three minutes for the user to press the SET REF button. If the user does 

not press the SET REF button within 3 minutes, the microcontroller will retrieve the stored reference levels from memory. 

The system will also warn the user with a steady yellow LED light when it’s using the retrieved reference levels after a 

power interruption. To clear the yellow warning light, we recommend you reconfirm the output RF levels by using the setup 

procedure.
During normal operation the alarm contacts are energized (i.e. the Normally Open contact is connected to the Common 

contact). This configuration ensures that the user be alerted when power goes down.

Summary of Contents for Q series

Page 1: ...INSTALLATION OPERATION MANUAL QPAIR Redundancy Switch Amplifier System www atxnetworks com www atxnetworks com Q SERIES Amplifiers ...

Page 2: ...lOn Series MAXNET HFC Enhance PCI Filters Q Series FiberLinx are registered trademarks of ATX in the United States and or other countries SMACSM is a service mark of ATX in the United States and or other countries Products or features contained herein may be covered by one or more U S or foreign patents Other non ATX product and company names in this manual are the property of their respective com...

Page 3: ...nual i TABLE OF CONTENTS Page 1 DESCRIPTION OF INTERFACES 1 1 2 DESCRIPTION OF OPERATION 2 1 3 SET UP PROCEDURE 3 1 4 SPECIFICATIONS 4 1 4 1 Amplifier Specifications 4 1 4 2 Switch Specifications 4 1 5 SERVICE SUPPORT 5 1 5 1 Contact ATX Networks 5 1 5 2 Warranty Information 5 1 ...

Page 4: ...ii Q Series Amplifiers QPAIR Redundant Amplifier System Installation Operation Manual This page left intentionally blank ...

Page 5: ...ected upon unplugging either power supply from the power source Anytime the PWR failure LED is lit and voltage has been restored or immediately after installation the RESET button must be pushed to activate normal operation RF Amplifier Connections Rear View a Rear of Chassis this section only describes the amplifier connections not the switch connections Input to AMP 1 labeled RF IN and located i...

Page 6: ...l level is approximately 25 dB lower than the signal level feeding the primary amplifier FROM AMP 1 Connect the output of the primary amplifier AMP1 to this port FROM AMP 2 Connect the output of the secondary amplifier AMP2 to this port RF OUT on left side of chassis and on the right side of the port labeled FROM AMP 2 This is the main output of the QPAIR system Installation Description 1 If you a...

Page 7: ...ence levels Normal on power up or reset Flashing 3 sec fast switched to the secondary amplifier when pushing and holding the FORCED SW button Red LED Failure On continuous primary amp failure Signal has been switched to the secondary amplifier by the system Flashing 1 sec slow secondary amp failure Flashing 3 sec fast input signal low or primary amp and secondary amp failure Red LED PWR Failure On...

Page 8: ...1 4 Q Series Amplifiers QPAIR Redundant Amplifier System Installation Operation Manual CHAPTER 1 DESCRIPTION OF INTERFACES This page left intentionally blank ...

Page 9: ...eration push the RESET button and then push the SET REF button If the system is running on the primary amp and the secondary amp level is down 3 dB or more the system considers the secondary amp failed and will set the alarm and set the red LED flashing slowly You need to troubleshoot or replace the secondary amplifier See Appendix 3 for troubleshooting instructions or Appendix 2 for replacing the...

Page 10: ...2 2 Q Series Amplifiers QPAIR Redundant Amplifier System Installation Operation Manual CHAPTER 2 DESCRIPTION OF OPERATION This page left intentionally blank ...

Page 11: ... chassis 11 Insert Change the proper pad and equalizer so the output levels of the amplifier AMP 1 are correct 12 Push the SET REF button 13 Connect signal strength meter to the OUTPUT TEST POINT of AMP 2 on the front of the chassis 14 Push and hold the FORCED SW button to force a switch to the secondary amplifier AMP 2 15 Insert Change the proper pad and equalizer so the output levels of the ampl...

Page 12: ...ads on the standoff of the circuit board These screws are located in the center of the board Do not remove the Phillips screws that hold the hybrid module in place Only loosen the 2 knurled thumbscrews in the horizontal centerline of the board Now you should be able to pull up on the screws to free the board from the push on connectors You will then need to tilt the board as you remove it so as to...

Page 13: ...screw is located on the front of the power supply The screws on the top and bottom are located on the tab of the power supply enclosure where it fastens to the chassis The front screw goes through the front plate of the QPAIR Just remove these 5 screws and then hold the power supply as you locate the 2 wires connecting to the chassis Just unplug the wires from their sockets and the power supply re...

Page 14: ...ets and Studs on Power Supply Connection Figure 6 QPAIR Block Diagram QPAIR Redundant Amplifier System TO AMP 1 RF INPUT TO AMP 2 FROM AMP 1 RF OUTPUT FROM AMP 2 AMP 1 IN AMP 1 OUT AMP 2 IN AMP 2 OUT NC NO NC NO PAD EQ PAD EQ PAD EQ PAD EQ 20 dB OUTPUT T P 20 dB OUTPUT T P AMP 1 AMP 2 RF SENSING SWITCH DETECTOR DETECTOR CONTROL ...

Page 15: ...20 1 16 260 40 0 78 75 4 PP GaAs QPAIR1000 19GP 40 1000 19 1 1 5 5 Pad EQ 20 1 16 375 40 0 77 73 5 GaAs PD QPAIR1000 23GP 40 1000 23 1 1 5 75 Pad EQ 20 1 16 375 40 0 77 73 4 GaAs PD QPAIR870 19GPH 40 870 19 1 1 5 5 Pad EQ 20 1 16 440 40 0 82 77 5 GaAs PDH QPAIR1000 19GPH 40 1000 19 1 1 5 5 Pad EQ 20 1 16 440 40 0 81 75 5 GaAs PDH QPAIR1000 23GPH 40 1000 23 1 1 5 75 Pad EQ 20 1 16 440 40 0 81 75 4 ...

Page 16: ...4 2 Q Series Amplifiers QPAIR Redundant Amplifier System Installation Operation Manual CHAPTER 4 SPECIFICATIONS This page left intentionally blank ...

Page 17: ... factory RF Products MAXNET SignalOn HFC Enhance PCI Filters Q Series SCN SMAC FiberLinx TECHNICAL SUPPORT Tel 905 428 6068 press 3 then press 2 Toll Free 800 565 7488 press 3 then press 2 USA Canada only Email rfsupport atxnetworks com CUSTOMER SERVICE ATX Networks 1 501 Clements Road West Ajax ON L1S 7H4 Canada Tel 905 428 6068 press 1 Toll Free 800 565 7488 press 1 USA Canada only Fax 905 427 1...

Page 18: ... ON L1S 7H4 Canada Tel 905 428 6068 Toll Free 800 565 7488 support atxnetworks com ISO 9001 15 REGISTERED 2018 ATX Networks Printed in Canada Information in this document is subject to change without notice www atxnetworks com Rev 07 18 ANW0837 ...

Reviews: