ATS 535-H Manual Download Page 3

INTRODUCTION 

Quality assurance tissue-mimicking phantoms are used to evaluate the accuracy and performance of ultrasound 

imaging systems. The phantoms mimic the acoustic properties of human tissue and provide test structures within 

the simulated environment. They are essential to detect the performance changes that occur through normal 

aging and deterioration of system components. Routine equipment performance monitoring can reduce the 

number of repeat examinations, the duration of examinations and maintenance time. 

This phantom is constructed of a new rubber-based tissue-mimicking material developed by ATS Laboratories. 

This material extends the useful life of the phantom by avoiding problems due to melting, freezing, dehydration 

and breakage from dropping which are common with hydrogel (water-based) phantoms. By eliminating these 

problems, the durability, quality and reliability of this product is guaranteed for three years. 

The acoustic properties of all biologic and non-biologic materials are affected by temperature variations.  Most 

diagnostic imaging systems and tissue-mimicking phantoms are calibrated at room temperature, commonly 

referred to as 23°C. To ensure measurement accuracy ATS incorporates a thermometer strip affixed to the outside 

surface of the phantom housing. 

The sound velocity of most diagnostic imaging systems is calibrated to 1,540 meters per second (mps), the 

assumed average velocity of sound through human soft tissue. The rubber-based tissue-mimicking material has a 

sound velocity of 1450 mps at 0.5dB/cm/Mhz at room temperature (23°C). The line targets and anechoic target 

structures have been physically positioned to compensate for the differences in the speed of sound, assuring 

accuracy of measurements.  

PRODUCT DESCRIPTION 

The Model 535-H rubber-based, tissue-mimicking (TM) phantom is designed to fulfill the basic testing 

requirements of a Quality Assurance Program.  

The phantom is designed with a combination of monofilament line targets and six tissue mimicking cylindrical 

targets of varying sizes. The monofilament line targets have a diameter of 0.12 mm, to optimize the displayed 

image at frequencies ranging from 2.25 to 7.5 MHz. Four groups of line targets are provided to evaluate the 

vertical and horizontal calibration measurements, axial-lateral resolution and the dead zone. 

TESTS PERFORMED 

·

 

Dead Zone or Transducer Ring-Down 

 

 

·

 

Vertical Measurement Calibration 

·

 

Horizontal Measurement Calibration  

·

 

Focal Zone 

·

 

Sensitivity 

·

 

Axial & Lateral Resolution 

·

 

Functional Resolution,  Definition and Fill-In 

Summary of Contents for 535-H

Page 1: ...MODEL 535 H BASIC QUALITY ASSURANCE PHANTOM Revised June 2014 ATS Laboratories Incorporated 900 Asbury Ave Norfolk VA 23513 Tel 800 617 1177 Email admin cirsinc com Website atslaboratories phantoms com ...

Page 2: ...Diagram 5 Dead Zone 6 Vertical Measurement Calibration 6 Horizontal Measurement Calibration 7 Axial and Lateral Resolution 8 Focal Zone 8 Sensitivity Maximum Depth of Penetration 9 Functional Resolution Definition and Fill in 10 Care and Handling of Rubber Based Phantoms 11 Warranty 11 Sample Ultrasound Performance Record ...

Page 3: ...To ensure measurement accuracy ATS incorporates a thermometer strip affixed to the outside surface of the phantom housing The sound velocity of most diagnostic imaging systems is calibrated to 1 540 meters per second mps the assumed average velocity of sound through human soft tissue The rubber based tissue mimicking material has a sound velocity of 1450 mps at 0 5dB cm Mhz at room temperature 23 ...

Page 4: ...Scan Surface 1 Scan Surface 3 Scan Surface 2 Scan Surface 4 ...

Page 5: ...umber of Targets 17 Interval Spacing 1 cm Depth 1 17 cm Horizontal Group Number of Targets 10 Number of Groups 2 Interval Spacing 2 cm Depth 5 cm from Scanning Surface 1 5 cm from Scanning Surface 3 Dead Zone Group Number of Targets 9 Lateral Displacement 5 mm Interval Spacing 1 mm Depth 2 10 mm Axial Lateral Resolution Group Number of Targets 6 Lateral Displacement 1 0 mm Interval Spacing 5 4 3 2...

Page 6: ... depth of the dead zone 7 Document the depth measurement on the quality assurance record Results If the depth of the dead zone is greater than 10 0 mm corrective action should be considered by the individual Ultrasound Department VERTICAL MEASUREMENT CALIBRATION Description and Reason For Testing Vertical distance measurements are obtained along the axis of the sound beam Proper diagnosis is depen...

Page 7: ...ration Linear array scanning systems should use 1 scanning surfaces Sector scanning systems should use 3 scanning surfaces 1 Place the phantom on a clean flat surface 2 Apply an adequate amount of low viscosity gel or water to the scan surface If water is used fill the scanning well slowly to avoid introduction of air bubbles 3 Adjust the instrument settings TGC output etc to establish baseline va...

Page 8: ...l liver scanning If the bottom of the phantom is seen adjust the gain settings until image goes entirely black Record these settings on the quality assurance record These settings should be used for subsequent testing 4 Position the transducer over the axial resolution group of line targets on the phantom until a clear image is obtained Freeze this image 5 Examine the image to determine if all of ...

Page 9: ...is the focal zone Measure the width of the beam and the depth at this point 7 Document the depth of the focal zone and the measurement of the focal width on the quality assurance record Results The location of the focal zone should agree with the manufacturer s specifications and should not change with time This applies to both fixed and dynamically focused systems If changes occur corrective acti...

Page 10: ...ystem is capable of resolving at a given depth Definition and Fill in describes the imaging system s ability to detect and display the shape and echogenic characteristics of a structure Clinically a correct diagnosis is dependent upon the system s ability to differentiate between a cystic or solid structure versus echo patterns originating from the surrounding normal tissue Testing Procedure The M...

Page 11: ...oided since they may adversely react with the rubber based material WARRANTY Statement of Warranty ATS Laboratories Incorporated warrants this rubber based phantom for it s lifetime from the date of delivery to the purchaser that the Phantom is free from functional defects in materials and workmanship The lifetime of this phantom is estimated to be 10 years from the date of manufacturing If ATS La...

Page 12: ......

Reviews: