4.
Ensure each pneumatic connection is fully seated by listening for an audible
“click” as each connector engages its receptacle. Gently tug on each tube to
ensure proper engagement with connector.
5.
Insert the corresponding red and black colored connections into the
thermocouple connectors, see Figure 15.
Figure 14
Figure 15
6.
The
cryo
ICE Probe icon, seen above in
Figure 14
, will extinguish if the
cryo
ICE
Probe is functioning properly and the approximate room temperature will be
displayed on the temperature display (typically 10 to 25° C). An example of this is
shown in Figure 16.
Figure 16: Probe Temperature Display
7.
A test run is advised to ensure the
cryo
ICE Probe and system is working properly
prior to the case.
8.
Pneumatic connectors should only be unplugged when the ACM is in the ready
mode.
xxii