Electromagnetic Immunity – Input A.C. Power Port
Guidance and manufacturer’s declaration – Input A.C Power Port Immunity
The
AtriCure
cryo
ICE BOX
is intended for use in the electromagnetic environment specified
below. The customer or the user of the
AtriCure
cryo
ICE BOX
unit should assure that it is used
in such an environment.
Phenomenon
Basic EMC standard or test
method
Immunity Test Levels
Professional healthcare facility
environment
Electrical fast transients /
bursts
a) l) o)
IEC 61000-4-4
± 2 kV
100 kHz repetition frequency
Surges
a) b) j) o)
Line-to-line
IEC 61000-4-5
± 0.5 kV, ± 1 kV
Surges
a) b) j) k) o)
Line-to-ground
IEC 61000-4-5
± 0.5 kV, ± 1 kV, ± 2 kV
Conducted disturbances
Induced by RF fields
c) d) o)
IEC 61000-4-6
3 V/m
m)
0.15 MHz - 80 MHz
6 V/m
m)
in ISM bands between
0.15 MHz and 80 MHz
n)
80% AM at 1kHz
e)
Voltage dips
f) p) r)
IEC 61000-4-11
0% U
T
; 0.5 cycle
g)
At 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
q)
0% U
T
: 1 cycle
And
70% U
T
: 25/30 cycles
h)
Single phase: at 0°
Voltage interruptions
f) i) o) r)
IEC 61000-4-11
0% U
T
: 250/300 cycle
h)