Troubleshooting
76
y
“Humidity sensor condensing.”
NO
NO
NO
YES
YES
YES
Condensation has formed on the
humidity sensor.
The cable of the sensor module is
connected properly to the sensor
module connecting port.
Dry the humidity sensor.
Connect the connector of the cable
of the sensor module securely.
The alarm message is still
displayed even after the sensor
module is replaced with a functional
one.
Failure of the sensor module
Failure of the detection board
❋
Condensation may form inside the sensor module when it gets cooled rapidly after the incubator has been
operated with a high humidity setting. The sensor module will return to its normal operation when the
condensation disappears.
Summary of Contents for Incu i 101
Page 2: ...Via F Croce 65 20081 Abbiategrasso MI Italy Tel 39 02 99763101 Fax 39 02 99763110 EU OFFICE...
Page 148: ...Parts List 146 9 1 Fig 1 Main Body and Accessories Fig 1...
Page 150: ...Parts List 148 9 2 Fig 2 Control Panel Fig 2...
Page 152: ...Parts List 150 9 3 Fig 3 Humidity Chamber Fig 3...
Page 154: ...Parts List 152 9 4 Fig 4 Mattress Platform Fig 4 The unit with the weight monitor...
Page 156: ...Parts List 154 9 5 Fig 5 Upper Section of the Main Body 1 Fig 5...
Page 158: ...Parts List 156 9 6 Fig 6 Upper Section of the Main Body 2 Fig 6...
Page 160: ...Parts List 158 9 7 Fig 7 Upper Section of the Main Body 3 Fig 7...
Page 162: ...Parts List 160 9 8 Fig 8 Lower Section of the Main Body Fig 8...
Page 164: ...Parts List 162 9 9 Fig 9 Power Unit Fig 9...
Page 166: ...Parts List 164 9 10 Fig 10 Oxygen Controller Assembly Fig 10...
Page 170: ...Parts List 168 9 12 Fig 12 Stand Fig 12...
Page 172: ...Parts List 170 9 13 Fig 13 Pulse Oximeter Masimo Fig 13...
Page 174: ...Parts List 172 9 14 Fig 14 Pulse Oximeter Nellcor Fig 14...
Page 186: ......
Page 188: ...C65SB100 3 18 15 Hongo Bunkyo ku Tokyo Japan Tel 81 3 38152311 Fax 81 3 38123144...