27
9.0 Technical Data
Mains voltage
100-240 V~ ± 10 %; 50/60 Hz
Power consumption
max. 45 VA
Fuses
2 x T 3,15 A / H 250V
Power cycle
Continuous operation
Light intensity
F 200
min. 120 klux
F 250
min. 80 klux
F 300
min. 55 klux
F 400
min 30 klux
Colour temperature
5000 ± 500 K
Cooling
Fanless / passive
Protective earth conductor resistance
max. 0,1 Ω
earth leakage current
max. 5 mA
Enclosure leakage current
max. 0,1 mA
Patient leakage current
max. 0,1 mA
Environmental conditions: Transport/storage
•
Temperature range
−10...+50°C
•
Air humidity without condensation
30...95 %
•
Air pressure
500...1060 hPa
Environmental conditions: Operation:
•
Temperature range
+10...+35°C
•
Air humidity without condensation
30...95 %
•
Air pressure
700...1060 hPa
Maximum operating altitude
≤ 3000 m
Contamination level
2
Overvoltage category
II
Weight:
3,65 – 5,6 kg
Periodical tests:
Repeat test of the electrical safety every 24 months.
Protection class against electric shock (acc. to EN
60601-1)
Recommended: inspection according to the manufacturer’s
specifications.
Degree of protection against electric shock
I
Degree of protection against Ingress
No applied parts
CE mark
IP X0
Ident-Nr. (REF)
538.0000.0 ATMOS i View 21
539.0000.0 ATMOS i View 31
538.9000.0 ATMOS i View 21 PRO
539.9000.0 ATMOS i View 31 PRO
605.0000.0 ATMOS i View 21 COLPO
606.0000.0 ATMOS i View 31 COLPO
Status of the Technical Data: June 2021