Calibration
The EZO
TM
ORP circuit can be calibrated
to any known ORP value
Cal,n
<cr>
calibrates the ORP circuit to a set value
<cr>
<cr>
Cal,clear
Cal,?
delete calibration data
device calibrated?
Command syntax
<cr>
<cr>
*OK
<cr>
*OK
Cal,225
<cr>
Cal,clear
Example
Response
<cr>
<cr>
?Cal,0
or
?Cal,1
*OK
<cr>
209.6 mV
225 mV
<cr>
Cal,?
Cal,225
<cr>