Replacing the Desiccant

Your  camera  includes  a  high-performance  molecular sieve  desiccant  tablet  which  is  used  to  

avoid condensation   inside   the   CMOS  chamber.   Although   this desiccant will last a long time 

it may eventually need replacing or recharging, in which case please follow the instructions below:

 

To replace the desiccant, open the desiccant port with the supplied plastic tool and a 

screwdriver and take out the used tablet.

To recharge the desiccant, place the tablet in an electric oven at 200ºC for 2 hours. Take the 

tablet out the oven (CAUTION: the tablet will be very hot) and allow it to cool down.

Place the recharged tablet back in the camera, replace the port cover and tighten (no need 

to over-tighten). Wait 24 hours before connecting the camera again.

We recommend leaving the camera in a warm, dry environment with the port open while you are 

recharging the tablets to ensure any remaining moisture can dry out.

You will notice that there is a filter inside the desiccant chamber. The purpose of this is to avoid 

contamination of the chamber when the desiccant is being replaced. The filter is very fragile and 

should not be touched.

Technical Specifications

Horizon

Image sensor

4/3 CMOS

Resolution

4644 x 3506

Pixel size (µm)

3.8x3.8

ADC

12bit

Readout noise (Typ.)

~1e- at x30 gain

Cooling Delta

-40°C

Regulated cooler

Yes

Water assist

No

Max. exposure

Unlimited

Min. exposure

18μs

Gain Factor

Variable 1 - x60

Full Well Capacity

~20,000e-

Dark Current

~0.016 e-/pix/sec at -10°C

PC Interface

USB2/USB3

Power requirements

12VDC, 2A

Backfocus distance

13mm ±0.5

Thread on front

M42x0,75

Weight

525g

Mono or Colour

Either

Summary of Contents for Horizon

Page 1: ...1 ATIK C A M E R A S Atik Horizon Camera Manual Version 1 0 November 2017 ...

Page 2: ... Window 5 Analog to Digital Converter ADC 5 Power Consumption 5 USB Port 6 Cable Post 7 Cooling 7 Dark Frames 7 Binning 8 Gain and Offset 8 Gain Presets 8 Offset for Advanced Settings 8 Replacing the Desiccant 9 Technical Specifications 9 Declaration of Conformity 10 Disposal of the Camera 11 Servicing and Repairs 11 Warranty 11 ...

Page 3: ...the most demanding astro imager in mind Your Atik camera incorporates state of the art design and materials and will be your trusted astro photography companion for a long time to come Further Information For easy installation instructions and useful information about how to take your first image please refer to the Quick Start Guide included with your camera Further information about our software...

Page 4: ... Your Atik Horizon box includes Atik Horizon CMOS Camera 2 Noserpiece and Cap USB3 Cable USB2 Cable Car lighter style power cable CD with software and manuals Quick Start Guide Cable Post and Cable Tie Desiccant Port Tool ...

Page 5: ...tion Atik Cameras are designed to have low power consumption to improve performance in the field Your camera is supplied with a power cable with a car light style plug which allows you to connection to most power tanks and battery packs available on the market today If you would like to run the camera from a 110 220V mains supply make sure it s of good quality with a 2 1mm type plug center positiv...

Page 6: ...o memory and low amp glow The typeof USB 2 or 3 connection dtermines the speed of transfer between this buffer to the PC For high speed you can use USB 3 0 or for a lighter more flexible cable use USB 2 0 The type of connection does not affect image quality The USB 3 0 cable is connected as shown below and can be connected to either a USB 3 0 or USB 2 0 port on your PC Please note that USB 3 0 tra...

Page 7: ...ature to image at will depend on your environment and we recommend using a repeatable stable temperature rather than simply the maximum the camera can reach on a given night After switching on the camera it s advisable to allow up to 5 minutes before taking images so that the temperature can stabilise At the end of an imaging session turn off the cooling and allow a few minutes before disconnectin...

Page 8: ... only 650e This means bright stars will saturate very quickly Although it may look bad due to the small full well depth the read noise will only be about 1 3 of what it was at x1 gain These kind of settings are very useful for very dim objects or narrow band imaging if you don t mind the bright objects saturating Gain Presets We provide three gain presets and the option to use your own custom sett...

Page 9: ...tighten Wait 24 hours before connecting the camera again We recommend leaving the camera in a warm dry environment with the port open while you are recharging the tablets to ensure any remaining moisture can dry out You will notice that there is a filter inside the desiccant chamber The purpose of this is to avoid contamination of the chamber when the desiccant is being replaced The filter is very...

Page 10: ...This product is not designed for any critical applications Critical applications means life support systems medical applications connections to medical devices commercial transportations nuclear facilities or systems or any other applications where product failure could lead to injury to persons or loss of life or catastrophic property damage This product is not a toy This is a class A product In ...

Page 11: ... no charge to the purchaser for parts or for labour if the fault is reported within the guarantee period provided however that Atik Cameras is able to duplicate the defect or problem at its facilities This warranty does not apply to damage that occurred as a result of abuse or misuse abnormal service or handling damage which may have been caused either directly or indirectly by another product or ...

Reviews: