ATI
Q45P pH System
Part 6
–System Maintenance
40
O&M Manual
Rev-B (7/15)
MESSAGE
DESCRIPTION
POSSIBLE CORRECTION
pH Cal Fail
Failure of pH calibration.
Clean sensor, get fresh cal solutions,
regenerate sensor (if necessary) and redo
calibration. If still failure, sensor slope may
be less than 80% or offset may be out of
range. Perform sensor tests as described
in sensor manual. Replace sensor if still
failure.
TC Cal Fail
Failure of temperature calibration.
Clean sensor, check cal solution
temperature and repeat sensor temp
calibration. TC calibration function only
allows adjustments of +/- 6 ºC. If still
failure, perform sensor tests as described
in sensor manual. Replace sensor if still
failure.
Eeprom Fail
Internal nonvolatile memory failure
System failure, consult factory.
Chcksum Fail
Internal software storage error.
System failure, consult factory.
Display Fail
Internal display driver fail.
System failure, consult factory.
mV Cal Fail
Failure of factory temperature
calibration.
Consult factory.