Model A15/82 Fluoride Monitor with Auto-Calibration
8
O&M Manual
Rev-G (6/10)
INSTALLATION
The main installation requirements for the ATI Fluoride Monitor are mechanical installation of the monitor
chemistry module, and sample inlet overflow chamber, connection of power to both monitor and chemistry
module, and connection of sample and drain piping to the sample overflow chamber supplied with the system.
Figure 1 on page 6 shows a schematic of a typical installation of a complete system mounted on the optional
stainless steel plate. The monitor must be located within 5 feet of the chemistry module due to the limited sensor
cable length. If the optional stainless steel mounting plate was purchased, all components mount to this plate,
with the plate mounted to a wall. If the stainless plate was not supplied, buffer and standard holders must also be
mounted.
Figure 2 provides details on the mechanical installation of the standard panel mount electronic assembly.
The panel cutout must be made accurately as the cutout tolerance is small. Panel mounting brackets are
included with the monitor and are attached as shown below.
6.10"
155mm
5.46"
139mm
.34"
8.6mm
3.54"
90mm
3.78"
96mm
SIDE VIEW
FRONT VIEW
3.78"
96mm
MOUNTING BRACKETS
(TYPICAL 2 PLACES)
MOUNTING BRACKET INSTALLATION
PANEL CUTOUT
3.60"
91.4mm
3.60"
91.4mm
Figure 2 - Panel Mount Dim & Installation