4.5 Low-Pass Filter
The power-on default selection is no filtering. The “filTC” field in
Section 8.4—Console “CAL” | “SET”
controls the current filter selection. The cutoff frequency (i.e. -3 dB frequency)
is dependent on the sample rate selection, which is defined in
. The cutoff
frequencies for the different sampling rates are listed in the following table.
Table 4.6—Low-Pass Filtering
Selected
Filter
-3dB Cutoff Frequency (in Hz)
at 488 Hz
Sample Rate
at 976 Hz
Sample Rate
at 1953 Hz
Sample Rate
at 3906 kHz
Sample Rate
at 7912 Hz
Sample Rate
0
200
350
500
1000
2000
1
58
115
235
460
935.10
2
22
45
90
180
364.04
3
10
21
43
84
169.52
4
5
10
20
40
81.24
5
2.5
5
10
20
39.84
6
1.3
3
5
10
20.31
7
0.6
1.2
2.4
4.7
9.37
8
0.3
0.7
1.4
2.7
5.47
Figure 4.1—Filter Attenuation at 0.5 kHz Sample Rate
-30.0 dB
-24.0 dB
-18.0 dB
-12.0 dB
-6.0 dB
0.0 dB
0 Hz
1 Hz
10 Hz
100 Hz
1000 Hz
0.3 Hz
0.6 Hz
1.3 Hz
2.5 Hz
5 Hz
10 Hz
22 Hz
60 Hz
200 Hz
Frequency
Attenuation