Fitting
Make sure that the correct o-ring seal is fitted to the inlet port to
prevent leakage. Screw the reservoir (clockwise) into the inlet port.
Note:
Demineralised water is used to test the pump at
manufacture. On shipment, there will be traces of
demineralised water in the system.
Filling
Unscrew the reservoir cover lock nut and remove the reservoir
cover.
Using clean, recommended fluid fill the reservoir to approximately
2/3 full.
Refit the reservoir cover and re-tighten the reservoir cover lock nut.
Connect the required pipes and equipment to the outlet ports.