9
Use a state-sanctioned disinfectant with proven
bactericidal, fungicidal, and virucidal properties.
The following disinfectants can be used in the United
States and Canada.
CaviCide
®
, CaviWipes
®
(manufactured by Metrex)
4) Remove the handpiece
from the motor and remove
the motor from the hose
(Fig. 12).
For reprocessing of the
handpiece, refer to the
Instructions For Use of the
handpiece’s manufacturer.
Fig. 12
Handpiece
Motor
Hose
5) Carry the motor to the
decontamination area.
7-2
Cleaning and Drying
Perform cleaning and drying using the manual or automatic
method.
Manual Method (Cleaning and Drying the Exterior)
Wash the motor under running water following the procedures
below.
(Water condition: ≤100°F (38°C), ≥1 gal/min (3.5L/min);
water should be of the same quality as drinking water)
1) Brush the exterior of
motors under running tap
water using a soft bristled
brush for one minute (Fig.
13).
Fig. 13
2) Rinse the motors under running tap water for 20 seconds.