Response time:
t90 < 7 min (temperature step 20 ºC)
Relative humidity:
Accuracy:
±3.0 %RH from 5 to 95 %RH at 23 °C
Measuring range:
0 to 100 %RH
Resolution:
0.1 %RH
Response time:
t90 < 30 s (humidity step 65 %RH, constant
temperature)
GENERAL
Power supply voltage:
9 to 30 Vdc
Current output
in case of error:
< 3.8 mA or > 22 mA
Recommended
calibration interval:
1 year
(relative humidity 1 year, temperature 2 years)
Protection:
housing with electronics IP65, sensors are
located in cover with IP40 protection
Storage temperature range: -30 to +80 °C
Storage humidity range:
5 to 95 %RH (no condensation)
Working position:
cable gland upwards
Electromagnetic
compatibility:
EN 61326-1
Weight:
approximately 150 g
Housing material:
ABS