Atal TRA-102ND-E Manual Download Page 1

 

 

 
 

 

 
 

TRA-102ND-E

 

 
 

Manual 

 
 
 
 

ATAL B.V. 

Ampèrestraat 35-37 

NL-1446 TR PURMEREND 

 

Postbus 783 

NL-1440 AT PURMEREND 

 

T

 (+31) 0299 630 610 

F

 (+31) 0299 630 611 

 

E

 

[email protected]

 

I

 www.atal.nl 

Summary of Contents for TRA-102ND-E

Page 1: ...TRA 102ND E Manual ATAL B V Ampèrestraat 35 37 NL 1446 TR PURMEREND Postbus 783 NL 1440 AT PURMEREND T 31 0299 630 610 F 31 0299 630 611 E info atal nl I www atal nl ...

Page 2: ... corners of the case and removing the lid Pass the connecting cable through released gland and connect the wires to terminals for device connection it is recommended to use a shielded cable with maximum length of 1200m and with external diameter of 4 to 8 mm Finally tighten the gland and screw the lid check the integrity of the seal The loop resistance value Rc Rm resistance of wires shall fulfill...

Page 3: ...s on The cables should be located as far as possible from potential interference sources Installation commissioning and maintenance may only be carried out by personnel with qualification by applicable regulations and standards DIMENSIONS MEASURED VALUES Temperature Accuracy 0 6 C Measuring range 30 to 80 C Resolution 0 1 C ...

Page 4: ...9 to 30 Vdc Current output in case of error 3 8 mA or 22 mA Recommended calibration interval 1 year relative humidity 1 year temperature 2 years Protection housing with electronics IP65 sensors are located in cover with IP40 protection Storage temperature range 30 to 80 C Storage humidity range 5 to 95 RH no condensation Working position cable gland upwards Electromagnetic compatibility EN 61326 1...

Page 5: ...ULATION OF THE MEASURED VALUES Temperature T C 6 875 x I mA 57 5 Relative humidity RH RH 6 25 x I mA 25 END OF OPERATION Disconnect the device and dispose it according to current legislation for dealing with electronic equipment TECHNICAL SUPPORT AND SERVICE Technical support and service is provided by distributor ...

Reviews: