30
11-3. Setting recorder output
Set the low and high-limit value of the recorder output range. Set for each scale as needed.
Table 11-3
Factory default setting
Setting range
The difference
between low and
high-limit value.
low-limit
value
high-limit
value
low-limit value
high-limit value
Refractive
index
1.31800
1.36600
1.31800 to 1.36590 1.31810 to 1.36600 0.00010 or more
Brix
-2.000%
21.000%
-2.000 to 20.990%
-1.990 to 21.000%
0.010% or more
Temperature
-35.0
℃
165.0
℃
-35.0 to 164.0
℃
-34.0 to 165.0
℃
1.0
℃
or more
①
Select the measurement scale (nD, Brix CH1 to 3, or
Conc). (Refer to pg. 22 "10. Measurement Scale Settings.")
The example below is for setting the alarm in Brix CH1
measurement mode.
②
Press the START/STOP key to pause the measurement.
③
Press the SCALE key 5 times.
The SET UP light will then blink, and "SEt" will be
displayed on the measurement value display (Fig.11-1).
④
Press the ENTER key and the SET UP light comes
on.Using the
↑
↓
keys, select "3. rEC" (Fig.11-10).
⑤
Press the ENTER key to enter the Recorder output
setting screen.
⑥
The temperature display will indicate "Lo", and the screen
for setting the recorder output low-limit value of refractive
index, Brix, or concentration will appear (Fig.11-11).
Use the
↑ ↓
keys to change the value, and select the
decimal place with the
←
→
keys.
Fig.11-10
Fig.11-11
Press the SCALE key to return to the previous
"3. rEC" screen. Thereafter, pressing the SCALE key
will prompt to return to the 3.rEC screen at any stage
during the recorder output setting and all of the previous
setting will be cancelled.