47
11-7. Measurement Interval Set-Up
●
Measurement interval: " c - # "
There are 5 options of intervals at which the unit takes
measurements. The factory default is the shortest
interval "c-1" wherein measurements are taken every
approx. 1second.
The other options indicate longer intervals; "c-2" is
twice as slow as "c-1," "c-3" three times as slow, and
so forth. The measurement interval setting is reflected
on the data transmission rate.
①
Press the START/STOP key to pause the measurement.
②
Press the SCALE key 5 times.The SET UP light will then
blink, and "SEt" will be displayed on the measurement
value display (Fig.11-1).
③
Press the ENTER key and the SET UP light comes
on.Using the
↑
↓
keys, select "7. C-no" (Fig.11-36).
④
Press the ENTER key. The measurement interval setting
screen will be displayed. Use the
↑
↓
keys to select
either C-1, C-2, C-3, C-4, or C-5. Press the
ENTER key to confirm the selection.
The display will return to the "7. C-no" screen.
⑤
Press the SCALE key once more to return to the
measurement results display screen.
Fig.11-36
※
Press the SCALE key to cancel setting procedures.
Examples of measurement interval
"c-1"
"c-5"
measurement
measurement
measurement
measurement
measurement
measurement
Approx. 5 sec.
Approx. 5 sec.
Approx. 1 sec.
Approx. 1 sec.