Offset Function
Automatic Temperature Compensation
Cleaning
Storage
This function modifies the measurement value by a constant. Use this function to match the measurement value to the refractometer (Brix meter).
ATAGO instruments are rigorously inspected to ensure each
unit meets the highest standards of quality assurance.
How to set addition/subtraction number
[Range:
-9.9~9.
9]
1. Hold down menu while
it is turned on.
2. Press menu to select
either addition or
subtraction.
(Example)
(Example)
No plus sign will be displayed.
Current addition/subtraction number is
displayed and
“-“ will flash.
4. Enter the addition/subtraction
number.
5.
Press START to confirm the
addition/subtraction number.
3. Press START
to confirm.
The first digit :
0 to 9
The second digit :
0 to 9
The first digit:
P r e s s m e n u t o
changethe number.
P r e s s S T A R T t o
confirm and move
to the next digit.
The second digit:
P r e s s m e n u t o
change the number.
The entered
addition/subtraction
number will flash first
then remains lit once
confirmed.
Press the START
button or side button
to begin measuring.
・Screen image during
measurement when
offset is on.
The measurement values are temperature dependent. This unit
is equipped with automatic temperature compensation to correct
the readings that are within the range to display measurement
value at 20℃.
The unit displays temperature at the fruit surface.
Safety Precautions
Repair and Warranty
Specifications
The instrument is warranted for one year from the date of
purchase. This warranty is void if the instrument shows
evidence of the following. Send the included batteries as well if
they are still in use.
・Having been disassembled by unauthorized personnel
・Water damage or having been dropped
・
Having been misused and/or operated outside the environmental
specifications
・
Leakage from batteries other than those included with the unit
Repair services are available for a fee after the warranty expires.
Contact an ATAGO authorized service center for service and
support.
・Do not aim it at a face when measuring.
・Do not attempt to repair, modify, or disassemble the instrument.
Brix scale
Brix represents the weight of sucrose in 100 grams of sucrose
solution as percentage by weight. When other dissolved solids
are present in the solution, Brix conversion may be applied.
Brix is a measure of the total dissolved solids in a solution and
indicates the combined concentration of all soluble substances,
such as sugar, salt, protein, and acids.
When the measured values are off
①Are there any difference between the fruit, unit, and ambient
temperature?
②
Are there any water droplets or soiled area in the measurement section?
Is the fruit spoiled?
③Does the fruit and inner cushion have a good contact?
Especially when using the side button, is it out of alignment?
④
Are there any difference depending on measured section?
①~④ → Refer to “
Measurement tips and precautions”
⑤
Are there any damages or soil on the light emitting or
receiving section? → Refer to “Cleaning”
If light emitting or receiving section is damaged or soiled,
contact ATAGO.
⑥ Is the cushion wet? → Refer to “Cleaning”
Replace cushion to a new one if there is any deterioration. →
Refer to “How to Change the Cushion”
⑦ Check the offset feature →
Refer to “Offset Function”
Measurement range
Resolution
Measurement accuracy
Repeatability
Automatic temperature
compensation range
Ambient temperature range
Measurement time
Backlight
Power supply
International Protection class
Dimensions and weight
*
Apple varieties and measurement environment may affect accuracy.
Brix 10.0 to 18.0%
Temperature 5.0 to 4
5.0℃
Measurement fruits
Apple
Brix 0.1% Temperature 0.1℃
Brix ±1%
* Temperature
±
1℃
Brix ±0.5%
5.0~35.0℃
5 to 35℃
Approx. 1 second
Two (2) AAA alkaline batteries
IP64
61 (W) x 44
(D)
x 115
(H)mm,
120g (main unit only)
2108K Printed in Japan
Gasoline
To clean the soiled sample stage, use a wetted tissue
or soft cloth to wipe the area.
Completely dry the cushion area.
Avoid using ethyl alcohol. It may compromise the
adhesion on the cushion.
Never splash
water on the
unit.
Do not use organic
s o l v e n t s ( p a i n t
thinner, benzene,
gasoline, etc.) on
the p l a s t i c b o d y
case.
Do not keep:
●
In a hot and humid place
●
Close to a heater
●
In direct sun light
●
Extreme low temperature
●
Inside a closed car
・Before use, carefully read this instruction manual and fully
understand the function and operation for each part of the
instrument. ・ATAGO is not liable for any or all damages as a
result of use of the instrument for those other than its intended
purposes (measurement of Brix, of fruits). ・Do not hit or
protrude the light emitting or receiving section with a sharp object.
・
ATAGO is not liable for any loss and damage caused by the
measurement and use of this instrument.
・Do not expose the
instrument to a rapid change in ambient temperature. ・Do not
use the instrument where there is much dust. ・Do not subject
the instrument to strong vibration. ・Do not place the
instrument under anything heavy. ・
If the instrument is dropped or
subjected to a strong impact,
c o n t a c t y o u r s u p p l i e r f o r
inspection. ・
Loosen the battery compartment cover for air
transportation.
Headquarters: The Front Tower Shiba Koen, 23rd Floor
2-6-3 Shiba-koen, Minato-ku, Tokyo 105-0011, Japan
TEL: 81-3-3431-1943 [email protected] http://www.atago.net/
TEL: 1-425-637-2107
TEL: 91-22-2833-8038 / 8076
TEL: 662-982-8718-9
TEL: 55 16 3916-6000
TEL: 39 02 36557267
TEL: 86-20-38108256
TEL: 7-812-777-96-96
TEL: 234-707-558-1552
menu
menu
(While the measurement is
displayed.)
WARNING
CAUTION
The backlight stays on for 30 seconds then
turns off after 30 seconds. The backlight is
off when measurements are taken.
TEL: 7-727-257-08-95 [email protected]
How to Change the Cushion
The Silicon cushion R, the Cushion RS and the Silicone Cover
can be replaced. Please contact ATAGO.
RE-39011 Silicon cushion R
1
.Pull off the Silicon cushion R from the sample stage.
2
.Fit the Silicon cushion R on to the sample stage.
RE-39012 Cushion RS
(
3sets)
1
.Remove the old cushion from the sample stage.
2
.Peel the paper backing off the sticker. Place the cushion on
the sample stage.
Do not block the radiation thermometer with the cushion.
RE-39415
JORDAN for PAL-HIKARi(Silicon Cover)
Note
1. Press and hold menu then press the
side
button while it is turned on
.
3.
While pressing menu, use the side button
to select ON/OFF.
2. Release the side button while
continuing to press menu.
Press START or
side button to
begin measuring.
Do not press menu for 5 seconds or more before
pressing the side button. Holding menu will activate
the addition/subtraction number setting.
Note
4. Release finger from menu.
Confirm setting.
Pressing the side button will switch it between ON/OFF.
(Example)
(Example)
Side button
ON/OFF will flash first then remains lit
once confirmed.
* Keep menu
pressed until step
3 is completed.
How to Set Alarm Sound and Alarm Vibrationcan
Buzzer sound and Vibration can be set to alarm when measurement error is displayed or at the beginning and end of measurement.
Buzzer
sound
ON/OFF
Buzzer
sound
Vibration
ON/OFF
Vibration
Snd
Snd
Snd
Snd
vbr
vbr
vbr
vbr
Switch it between Sound or Vibration
Press and hold menu then press the
start button.
Pressing the start button will switch it
between Buzzer Sound or Vibrationcan.
Maximum number of data history
Output
100
NFC Forum Type 4 Tag
ISO/IEC 14443 Type A
Output category:Date Time, Brix [%], Temp [degC]
(e.g.) 2017/08/17 09:30:45, 12.3, 20.4
For default Press the side button while setting up the Off-Set feature.