P-OQ0612-5291-E03
1/7
Operational
Q u alification (O Q
)
Automatic Polarimeter
AP-300
Operational Qualification of Automatic Polarimeter AP-300
Preface
The Operational Qualification performance test of the Automatic Polarimeter AP-300 is described
herewith.
The instructions below assume that the performance test of the AP-300 is conducted by a regular
user of the unit who is familiar with the operation. For daily check and maintenance, please refer
to the Standard Operational Procedure document (SOP).
1. Test Environment and Preparation
Conduct the performance test in an environment which meets the conditions outlined below.
Keep the measurement sample and the AP-300 at an ambient temperature of 18.0°C to 22.0°C
for at least 30 minutes to let them acclimate to the ambient temperature.
If the following environmental conditions are not met, an appropriate measurement result
may not be obtained.
Environment for performance test
2. Establishing a Discretionary Measurement Standard
A performance test requires a discretionary measurement standard.
Establish a set measurement standard for your samples, taking the accuracy of the AP-300
into account. Please refer to the product literature and/ or the instruction manual for the
specifications, features, and measurement accuracy of the AP-300. These specifications are
guaranteed at the time the AP-300 is shipped as a brand-new product. It is difficult to obtain
an accuracy level greater than the product specifications under different environmental
conditions. It is recommended setting the discretionary measurement standard lower than the
stated specifications.
Example: Judgment standard measurement
Accuracy: Angle of rotation ±0.03°
3. Confirmation of Measurement Accuracy (Accuracy and Repeatability)
Check the accuracy (accuracy and repeatability) of the measurement values by measuring a
sample of a known concentration.
The AP-300 is an optical instrument designed to measure the angle of rotation of a sample
solution by utilizing two polarizing plates.
Ambient temperature: 18.0°C to 22.0°C Relative humidity: less than 75%
Power supply: AC 100 to 240V, 50/ 60Hz
※
The Automatic Polarimeter AP-300 should be used indoors and placed on a flat,
level surface, such as a work desk or laboratory table.