-14-
6. Error Messages
The battery power is low.
Not enough light could pass through
the sample solution to take the angle
of rotation measurement.
a
Dilute or filter the sample before
measuring.
An error has occurred while
adjusting the setting.
Too much external light is interfering
with the measurement.
a
Shade the sample stage with
your
A
hand and measure again.
*
This error message may also
appear if the sample temperature
is too low.
The sample measured above
the measurement range.
Measurement error.
(When User scale is selected)
a
Press
J
and
K
to
A
view the detail of the error.
The sample measured below
the measurement range.
(Example)
The sample measured below 3.9˚C
(16.9˚C in the Purity or ISS scale).
G
was pressed with
nothing or something other
than water on the sample
stage. (When Brix or User
scale is selected)
(Example)
The sample measured above
41.1˚C.
■
-
3.0
.
0
..
HHH
℃
-
A
.
A
.
A
20.3
℃
■
-
3.0
.
0
..
LLL
℃
-
L
.
L
.
L
20.3
℃
-
E
.
r
.
r
..
bAS
-
H
.
H
.
H
20.3
℃
-
E
.
r
.
r
..LLL
-
n
.
nn
.
-
O
.
d
.
..LLL
-
L
.
o
.A
℃