Zero-setting  and  Measurement 

 

Zero-setting 

【Caution】 

 

Zero-set the instrument at the beginning of each day before use as well as after replacing the batteries. 

 

Let water on the prism acclimate to the temperature of the instrument before zero-setting. 

 

When 

AAA

 is displayed, wipe the prism clean, apply water, and press the ZERO button again.

 

 

1.

 

Wipe the prism clean and 

apply about 0.3mL of tap water. 

 

 

2.

 

Press the START 

button, and after 

---

 

appears, the 
measurement and the 
prism temperature are 
displayed. 

 

 

3.

 

 

 

4.

 

Zero-setting is 

complete when 

000

 blinks 

twice and then stops. 

 

Measurement : not 0.0 

Press the ZERO 
button while the 
water is on the 
prism. 

 

 

Measurement : 0.0 

Proceed to measurement. 
Wipe the prism clean with 
tissue paper. 

The instrument is zero-set 
correctly. 

 

Dilution 

 

<Reference> 

Ethyl alcohol

  76.9-81.4vol% 

The 1:1 dilution 

Apply some diluted 

sample 

 

 

 
←Water、100mL 

99%

Ethyl alcohol

  80mL

 

 
 

 

 

Water  1mL             

Ethyl alcohol 

1mL 

 
 
Gently stir. 

 

 

 

 
←Water、100mL 

95%

Ethyl alcohol

  83mL

 

 

Measurement 

 

1.

 

Wipe the prism clean and 

apply about 0.3mL of the sample. 

 

 

2.

 

Press the START 

button 
 

 

 

 

3.

 

After 

---

 appears, the 

measurement and the prism 
temperature are displayed. 
 

 

 

   

 

Displayed 
=measured value×2
(dilution) 
 
Offset 
y=ax+b 

Default:

  a=2  ,  b=0

 

 

【Caution】 

 

This instrument displays the concentration of ethanol in water as a percentage of total volume. 

 

When a solution contains components other than ethanol and water, the readings may not correspond to alcohol by volume. 

 

Do not use metal tools to apply samples on the prism as they may scratch the prism. 

 

Initial measurements may fluctuate with hot or cold samples. Wait for the instrument to acclimate to the sample temperature, 
approximately 20 seconds, to press the START button. Alternatively, press the START button multiple times until 
measurements become stable.

 

 

Do not splash water above 50℃. The plastic may warp, which may compromise the water resistance. 

 

The displayed temperature is that of the prism and may not necessarily match the temperature of the sample. 

 

<LCD Auto Shut-off> 

The instrument will turn itself off after 2 minutes of inactivity. To manually turn it off, hold down the START 
button for more than 2 seconds. 

<Option> 

An anti-volatile cover, MAGIC, is available (sold separately) to prevent volatilization of the sample. 
Part No. RE-39446      Apply some sample on the prism, then place the MAGIC so it fits tightly over the sample. 
Press the START key whilst leaving the MAGIC on the sample. 

 

 

Cleaning 

 

【Caution】 

 

Do not scratch the prism. 

 

The instrument is water-resistant, not waterproof, and should not be submerged. 

1. Wipe off the sample. 
2. Clean the prism and sample stage using a mild soap and thoroughly rinse with water.     
3. Dry the area with tissues thoroughly. 

 

Offset Function 

 

 

1.

 

Hold down 

ZERO while it 
is turned on. 

b

 wil appear. 

 

2.

 

Press ZERO 

to select either 
addition (b) or 
subtraction (-b). 

 

 

3.

 

Press START 

to confirm. 
 

 

4.

 

Enter the addition/subtraction number. 

ZERO to change the number: 

 

0

1

2

,  …, 

8

9

10

0

1

2

… 

START to confirm and move to the next 
decimal place. 
When the 1

st

 place is confirmed with 

10

,

 

the number selections for the decimal 
places are skipped. 

 

5.

 

Press START 

to confirm the 
addition/subtracti
on number. 
Next is to program 
a coefficient. 

(Factory default 
2.00)

 

 

6.

 

Enter the 

coefficient. 
 
Refer to the 
step 4. 
 
 
 

 

7.

 

Press 

START to 
confirm the 
coefficient. 

 

 

To disable the Offset feature, set the Offset value to factory default value 
(a:1.00, b:0.00). 

 

The measurement range is shifted according to the offset settings. 

 

Screen images when offset is on 

 

y  =  ax  +  b 

y:Actual concentration 

 

x:PAL-102S readings 

 

a:Coefficient (multiplication) 

 

  Range: 0.01 to 10.00

(Factory default 2.00)

 

 

b:Addition/subtraction number 

 

  Range: -10.00 to 10.00(Factory default 0.00) 

 

Replacing  the  Batteries 

 

【Caution】 

 

Fasten the battery compartment cover tightly to prevent water ingress or poor connection, which will cause erroneous 
measurements. Push the cover in firmly and turn.   

 

When the O-ring on the cover is dirty or damaged, the water resistance may be compromised. (Figure 1)   

 

When the battery icon indicates the low power level (

), replace both batteries with a brand new set of AAA alkaline 

batteries (1.5V).   

 

Static images may occasionally appear on LCD. Such retained pixel charges do not indicate a faulty display, consume the 
battery power, or affect the instrument

s performance in any way. 

 

Check the expiration dates on batteries before purchase.   

 

Zero-set the instrument after the batteries are replaced. 

 

1.

 

Insert a coin in the 

groove on the battery 
compartment cover. Turn 
the coin counterclockwise 
to remove the cover. 
 

 

 

2.

 

Insert batteries, 

observing the correct 
polarity. 
 

 

 

3.

 

Align the cover and 

push it down. 
 
 

 

 

4.

 

Close the battery compartment 

cover by pushing the cover in with 
a coin in the groove and turning it 
clockwise until it stops. 

*Turning excessively may cause 

malfunction.

 

 
 
 
 

 

1  2  3 

Figure1 

1 

O-ring 

Groove 

Approx.   
2mm 

Approx. 0.3mL of sample 

Approx. 0.3mL of water 

Approx.   
2mm 

For addition, only 

b

 will appear. 

1  2  3 

Water after 
zero-setting 

Offset 

b

 =   

addition of 0.30 

During 
measurement 

For coefficient (a) 

For addition/subtraction (b) 

No plus sign will 
be displayed. 

Mild 

soap 

MAGI

 

20.4

 

79.2 