Mounting procedure (Fig.
6-1, Fig. 6-2, and Fig. 6-3)
①
Install the CM-
BASEα so that the prism surface is at a right angle to the ground.
②
Attach the sample inlet unit to the CM-
BASEα with O-ring inserted between them, and
fasten them together with the clamp band.
③
Install the inlet unit so that the sample solution runs from the lower nozzle to the upper
nozzle to prevent air bubbles from forming.
④
When connecting the tubes to the hose connector, clamp them with a tie band. The tie
bands provided with the hose connector are made of plastic. If chemicals corrosive to
plastics are used, substitute with tie bands made of other material.
⑤
The prism surface may become contaminated with solids, dirt and/or grease. If this
happens, the prism surface must be cleaned by hand. (See page 22.) The sample
inlet unit should be installed in such a manner that it can be easily removed to allow
access to the prism for cleaning.
Fig. 6-1
Fig. 6-2
Fig. 6-3
Dimensions
Prism stage
O-ring (silicon
or EPDM)
Bottom
Top
Flow of solution
Flow of solution
Prism stage
O-ring
(silicon or EPDM)
Sample inlet
unit
(optional)
ex.hose connector
Clamp band