11. Setting Measurement SET UP Mode
(1) Calibration Procedure
①
Make sure that the detection section (sample inlet
unit) is properly mounted on the piping.
②
Let distilled water or a reference sample flow into
the piping.
※
If the last measurement value was LLLL.LL or
HHHH.HH, the screen for setting the bias will not
appear.
③
Turn on the power as instructed in "9.Turning On
the Power," on page 22.
④
Perform calibration in the appropriate measurement
scale for your sample (i.e. nD, Brix, or Conc1 to 5).
The example below is for calibration in Brix
measurement mode.
⑤
Press the START/STOP key to pause the
measurement.
⑥
Press the SCALE key three times.
The SET UP light will then blink, and "SEt" will be
displayed on the measurement value display (Fig.
※
If left inactive for 10 seconds, the display returns to
the "StoP" screen, following three short error beeps
Fig. 11-1
Fig. 11-2
⑦
Press the ENTER key.
The SET UP light comes on and "1. bIAS" will
appear on the measurement value display (Fig.
※
If left inactive for 10 seconds, the display returns to
the "StoP" screen, following three short error beeps
Press the SCALE key to return to the previous "SEt"
screen (Fig. 11-1) (During the SET UP operation, the
SCALE key serves as the cancel key).
Fig. 11-3