AL390
Highly Accurate Non-Contact Pyrometer With Analogue Output and Digital Interface
300°C to 1400°C
Accurate Sensors Technologies
Accurate Sensors Technologies
AL390
IR Pyrometer model provides the advantage of noncontact temperature
measurement of metallic surfaces and in flame heated furnaces i.e through
flames and flue gas. At a wavelength of 3.9um hot furnaces gases and clean
burning gas flames are quite transparent and therefore permit temperatures of
substances to be measured as well during heating up process in industrial
furnaces with the use of infrared Pyrometer
AL390
.This format combines the high
accuracy of digital signal processing with the simple connection.
AL390
IR
Pyrometer is used for target temperature ranging from 300°C to 1400°C. The
electronic assembly is protected by IP65 rugged stainless steel housing.
AL390
IR Pyrometer is provided with USB 2.0 for parameter setting & Data Logging.
Technical specifications
Ÿ
Highly Accurate due to digital linearization of the output
Ÿ
Temperature range from 300°C....1400°C
Ÿ
Spectral range 3.9 µm
Ÿ
Four wire form with Analog output 4...20mA or 0...20mA or
0...10V
Ÿ
Serial communication RS-232/RS-485
Ÿ
Response time 60 ms
Ÿ
Very good stability
Ÿ
Laser Targeting for precision targeting
Ÿ
User friendly Software
Features
Ÿ
Measurement Through Flames & Combustion Without
Influencing The Measurement.
Applications
Ÿ
Mechanical& Electrical Accessories
Ÿ
Extra Cable lengths
Optional
Ÿ
Pyrometer with Laser pilot light(PL)
Ÿ
Analog output 4...20mA, 0..20mA, 0..10V
Ÿ
Digital USB 2.0 output, USB Cable
Ÿ
Digital Interface RS-232 / RS-485
Ÿ
5 mtr long connection cable with connector
Ÿ
1.5 mtr. Communication Cable
Ÿ
Calibration certificate, Software & Operation manual
Standard Scope of Supply
Model
AL390
Temperature Range
(Analog sub-range
adjustable)
Spectral Range
Photodetector Type
Distance to Spot Size Ratio
Emissivity
(ε)
Response Time
Accuracy
Repeatability
Sighting Option
Analog Output
Digital Output
Operating Temp. Range
Storage Temp. Range
Adjustable Parameters and
Features via Software
300°C….1400°C
3.9 µm
Thermopile
50:1
0.1….1.2 adjustable
60 msec adjustable upto 10sec
T < 500°C ,± 1.5% of measured value
T
≥
500°C, ± 1% of measured value
(The instrument must be at constant ambient
temperature for minimum 25-30 mins in power on
condition)
0.3% of reading in °C +1°C
Laser Pilot Light (PL)
USB 2.0
RS-232/RS-485 (User selectable)
*At a time only one digital output possible
0°C…….70°C
0°C…..200°C (With water cooling jacket)
-20°C…70°C
Emissivity, Response Time, Clear Time(Peak Picker),
Analog Output, Analog Scale (Sub Range) , Unit Of
Temperature(°C/°F), Communication mode
(Comm.mode), Record feature etc
0-20mA,4-20mA,0-10V (User selectable)
Power Supply
Power Consumption
Laser Power
Protection Class
Housing
Isolation
Operating Humidity
Weight & Dimensions
24V DC
Max 2.5 watt
<1 m watt
IP65
Stainless Steel
Power supply,*Digital output and Analog output are
galvanically islolated against each other
* Not applicable for USB 2.0 digital output
10-95%, Non-Condensing Conditions
600g
Dia= Ø 49.5 mm; Length=118mm
We measure accurate
in extreme conditions
temperature