Associated Research HYPOTULTRA III 7620 Quick Start Manual Download Page 3

BACK

PANEL

CONTROLS

1

.

BUS INTERFACE

: Standard connector for interconnection to the USB/RS-232

Bus interface. Optional IEEE–488 interface or Ethernet Interface may be 
substituted for the USB/RS-232.

2

.

CALIBRATION BUT TON

: Used to put the instrument into calibration mode.

3

.

REMOTE SIGNAL INPUT

: 9-Pin D sub-miniature male connector for remote 

control of test, reset, and interlock functions, as well as remote program 

memory selection.

4

.

REMOTE SIGNAL OUTPUT

: 9-Pin D sub-miniature female connector for 

monitoring pass, fail, processing and reset out output relay signals.

5

.

SCANNER CONNECTOR

: For connection of optional External Scanning 

Matrix.

6

.

HIGH VOLTAGE OUTPUT TERMINAL

: Connector used to attach the high 

voltage test lead, adapter box high voltage lead, or test fixture high 

voltage lead to the instrument. The connector is recessed for safety.

7

.

RETURN OUTPUT TERMINAL

: Connector used to attach the return test lead or 

test fixture return lead to the instrument. This connection provides the 

return current path for the high voltage and continuity tests.

8

.

CONTINUIT Y OUTPUT TERMINAL

: Connection for the low current return lead 

or three-prong receptacle adapter box used for continuity testing.

9

.

FUSE RECEPTACLE

: Socket for the input power fuse. See operation and 

service manual for ratings and replacement instructions.

10

.

SCANNER OUTPUTS

: Optional scanner matrix that provides 8 HV, 

Continuity, or Return connections. Please refer to the options section of the 

product manual for additional connection information.

11

.

INPUT POWER RECEPTACLE

: Standard IEC 320 connector for connection 

to a standard NEMA style line power (mains) cord.

12

.

CHASSIS GROUND (EARTH) TERMINA L

: This terminal should be 

connected to a good earth ground before operation.

2

Summary of Contents for HYPOTULTRA III 7620

Page 1: ...nnections NEVER perform a test on energized circuitry or equipment BE SURE to always connect the return test lead first HANDLE test clips by insulation only never touch clips directly WARNING THIS GUIDE WAS CREATED FOR OPERATORS HAVING SOME FAMILIARITY WITH HIPOT AND ELECTRICAL SAFETY TESTING A HIPOT TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK TO PREVENT AC...

Page 2: ...ction provides the return current path for the high voltage 7 POWER SWITCH Rocker style power switch with International ON I and OFF O markings 8 SOFT KEYS Multi function parameter selection keys used to select screens and change parameters 9 EXIT KEY Key used to escape from parameter editing and returning to prior screens 10 ENTER KEY Key used to finalize parameter entries The ENTER key may also ...

Page 3: ...high voltage lead to the instrument The connector is recessed for safety 7 RETURN OUTPUT TERMINAL Connector used to attach the return test lead or test fixture return lead to the instrument This connection provides the return current path for the high voltage and continuity tests 8 CONTINUITY OUTPUT TERMINAL Connection for the low current return lead or three prong receptacle adapter box used for ...

Page 4: ...Turn the POWER switch to ON The initialization screen will appear After three seconds the Perform Test screen will appear See Figure 1 WARNING LOCATE A SUITABLE TESTING AREA WITH A THREE PRONG GROUNDED OUTLET BE SURE THAT YOUR THREE PRONG OUTLET HAS BEEN TESTED FOR PROPER WIRING ALSO MAKE SURE YOU READ THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST 3 EDITING TEST PARAMETERS FROM EDIT B...

Page 5: ...est screen Figure 2 Test Parameters Screen 4 SELECTINGATESTMEMORY From the Perform Test screen press the Memory soft key The Memory Recall screen will now be displayed There are two methods to select a memory Enter a Number or List and Scroll ENTER A NUMBER Enter the number with the numeric key pad that you would like to use Press the ENTER key to accept the new number or the EXIT key to escape fr...

Page 6: ... the Select soft key to choose the character Press ENTER after the memory name is complete The current test parameters will now be saved to this location and the Memory Recall screen will now be displayed 5 ADAPTER BOX CONNECTION The adapter box provides an easy way to connect the HypotULTRA III to a device under test DUT terminated in a two or three prong line cord The following diagram shows how...

Page 7: ...the HypotULTRA III to your DUT The Continuity function must be turned OFF for this connection See Figure 4 Plug one of the black ground return leads into the Return terminal on the front panel of the instrument Then connect the high voltage lead terminated in a red clip into the High Voltage terminal on the front panel of the instrument Attach the clip terminated end of the return lead to the dead...

Page 8: ... See Figure 5 SELECTING THE RESULTS SETTING From the Perform Test screen press the More soft key A second set of selections will now be displayed Press the System soft key Using the Arrow soft keys scroll to the results parameter and press the Edit soft key Use the Change soft key to select how the results will be displayed Press the ENTER key to save The available selections are All P F and Last ...

Reviews: