BACK
PANEL
CONTROLS
1
.
BUS INTERFACE
: Standard connector for interconnection to the USB/RS-232
Bus interface. Optional IEEE–488 interface or Ethernet Interface may be
substituted for the USB/RS-232.
2
.
CALIBRATION BUT TON
: Used to put the instrument into calibration mode.
3
.
REMOTE SIGNAL INPUT
: 9-Pin D sub-miniature male connector for remote
control of test, reset, and interlock functions, as well as remote program
memory selection.
4
.
REMOTE SIGNAL OUTPUT
: 9-Pin D sub-miniature female connector for
monitoring pass, fail, processing and reset out output relay signals.
5
.
SCANNER CONNECTOR
: For connection of optional External Scanning
Matrix.
6
.
HIGH VOLTAGE OUTPUT TERMINAL
: Connector used to attach the high
voltage test lead, adapter box high voltage lead, or test fixture high
voltage lead to the instrument. The connector is recessed for safety.
7
.
RETURN OUTPUT TERMINAL
: Connector used to attach the return test lead or
test fixture return lead to the instrument. This connection provides the
return current path for the high voltage and continuity tests.
8
.
CONTINUIT Y OUTPUT TERMINAL
: Connection for the low current return lead
or three-prong receptacle adapter box used for continuity testing.
9
.
FUSE RECEPTACLE
: Socket for the input power fuse. See operation and
service manual for ratings and replacement instructions.
10
.
SCANNER OUTPUTS
: Optional scanner matrix that provides 8 HV,
Continuity, or Return connections. Please refer to the options section of the
product manual for additional connection information.
11
.
INPUT POWER RECEPTACLE
: Standard IEC 320 connector for connection
to a standard NEMA style line power (mains) cord.
12
.
CHASSIS GROUND (EARTH) TERMINA L
: This terminal should be
connected to a good earth ground before operation.
2