Associated Research HYAMP III Quick Start Manual Download Page 2

EDIT SYSTEM PARAMETERS

  

Select the correct input line voltage on the rear panel of the instrument,  
either 115 VAC or 230 VAC.

  

Connect the female end of the power input plug into the rear of the 
instrument and plug the male end of the cord into a grounded power source.

  

Plug the Interlock Disable Key into the signal/input connector on the rear 
panel of the instrument.

  

Turn the POWER switch to ON.

The initialization screen will appear. After three seconds the Perform Test
screen will appear as shown below.

Note:

 3160 Owners: Please reference your product manual for rear panel controls.

Configure the instrument system parameters to your preferences. The instrument 
system parameters are global and will affect all tests that you perform regardless 
of memory location and memory step.

SIGNAL INPUT

CAL

SIGNAL OUTPUT

OPTION

CAT.II

T6.3A 250V

VOLTAGE

SELECTOR

INTERCONNECT

RETURN

CURRENT

GND

115V~/ 230V~

6.3A MAX., 50Hz/60Hz

11

5

REMOTE INTERFACE

LINE INPUT

SIGNAL INPUT

CAL

SIGNAL OUTPUT

OPTION

CAT.II

T6.3A 250V

VOLTAGE

SELECTOR

INTERCONNECT

RETURN

CURRENT

GND

115V~/ 230V~

6.3A MAX., 50Hz/60Hz

11

5

REMOTE INTERFACE

LINE INPUT

SIGNAL INPUT

CAL

SIGNAL OUTPUT

OPTION

CAT.II

T6.3A 250V

VOLTAGE

SELECTOR

INTERCONNECT

RETURN

CURRENT

GND

115V~/ 230V~

6.3A MAX., 50Hz/60Hz

11

5

REMOTE INTERFACE

LINE INPUT

3130

3130

3130

Perform Test Screen

WARNING: LOCATE A SUITABLE TESTING AREA WITH A THREE-PRONG, GROUNDED 
OUTLET. BE SURE THAT YOUR THREE-PRONG OUTLET HAS BEEN TESTED FOR PROPER 
WIRING. READ THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST.

INSTRUMENT SETUP

3140/3145

3140/3145

3140/3145

START

From the Perform Test screen, 
select the MENU soft key twice.

Select the SYSTEM soft key.

Select the “+” soft key to change 
the parameter.

GND   Settings
M1-1     1.0s

____________________________

Menu

A

M1-1   GND
  Settings

________________

    5.00A
     117mÍ
       0mÍ
     1.0s

Results

Test

System

Exit

PLC Remote   OFF
Single Step  ON
Alarm        5
Contrast     1
Results      Last
Lock         OFF
Mem Lock     ON
     <More>

 

 

+

Exit

>

>

PLC Remote   OFF
Single Step  ON
Alarm        5
Contrast     1
Results      Last
Lock         OFF
Mem Lock     ON
     <More>

 

 

+

Exit

>

>

Select a parameter to edit by 
scrolling with the       and      
soft keys.

Select EXIT twice to save changes
and return to the Perform Test screen.

PLC Remote   OFF
Single Step  ON
Alarm        6
Contrast     1
Results      Last
Lock         OFF
Mem Lock     ON
     <More>

 

 

+

Exit

>

>

To Grounded 

Power Source

2

3

With our PPE Safe  

Workstation Packages

KEEP YOUR 

OPERATOR SAFE

Scan for details

Summary of Contents for HYAMP III

Page 1: ...lips directly and always connect the return lead first You should always know when a test is being performed WARNING THISGUIDEWASCREATEDFOROPERATORSHAVINGSOMEFAMILIARITYWITHELECTRICALSAFETY TESTING AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK TO PREVENT ACCIDENTAL INJURY OR DEATH THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN HAN...

Page 2: ...OTE INTERFACE LINE INPUT SIGNAL INPUT CAL SIGNAL OUTPUT OPTION CAT II T6 3A 250V VOLTAGE SELECTOR INTERCONNECT RETURN CURRENT GND 115V 230V 6 3A MAX 50Hz 60Hz 115 REMOTE INTERFACE LINE INPUT 3130 3130 3130 Perform Test Screen WARNING LOCATE A SUITABLE TESTING AREA WITH A THREE PRONG GROUNDED OUTLET BE SURE THAT YOUR THREE PRONG OUTLET HAS BEEN TESTED FOR PROPER WIRING READ THE SAFETY CHECKLIST OF ...

Page 3: ...ove the leads push the lead in toward the instrument and then pull back on the lead to fully remove lead DID YOU KNOW Always connect the ground return clip first and double check that both clips have a solid connection to the DUT START From the Perform Test screen select the MENU soft key twice On the Test Settings screen select the TEST soft key Select EDIT once you have chosen the correct parame...

Page 4: ...nnect terminal on the HYAMP III Ground Bond tester Plug the other end into the Return terminal on the rear panel of the Hypot III Using the rear panel connections connect the following cords P N 5030DT 21 Plug the blue coded end into the Ground Bond interlock connector Plug the red coded end into Hipot signal output P N 5030DT 22 Plug the blue coded end into the Ground Bond signal output Plug the ...

Page 5: ...DEO For additional information about these and other key features of the HYAMP III please consult the full Operation and Service Manual or call us toll free 1 800 858 TEST 8378 or 1 847 367 4077 2017 Associated Research arisafety com 1 2017 FOLLOW US ...