P/N 38490 - Plug the black return test lead into the front panel Return terminal on
the HYAMP
®
.
Clip the end of the test lead to chassis ground of the DUT.
P/N 38482 - Plug the red high current lead of the adapter box into the front panel
Current terminal on the HYAMP
®
.
Plug the white Alden plug into the front panel High Voltage terminal on the Hipot
tester.
CONNECTION TO A DEVICE UNDER TEST
2
3
4
5
Plug the DUT line cord into the adapter box.
Once the test parameters are set, start the test from the HYAMP
®
.
If the Ground Bond test passes, the Hipot test will automatically start. If a
test fails, the test sequence will abort and failure results are displayed.