background image

7

P/N 38490 - Plug the black return test lead into the front panel Return terminal on 
the HYAMP

®

Clip the end of the test lead to chassis ground of the DUT.

P/N 38482 - Plug the red high current lead of the adapter box into the front panel 
Current terminal on the HYAMP

®

.

Plug the white Alden plug into the front panel High Voltage terminal on the Hipot 
tester.

CONNECTION TO A DEVICE UNDER TEST

2

3

4

5

  

Plug the DUT line cord into the adapter box.

  

Once the test parameters are set, start the test from the HYAMP

®

.

  

If the Ground Bond test passes, the Hipot test will automatically start. If a

 

 test fails, the test sequence will abort and failure results are displayed.

RETURN

CURRENT

RESET

TEST

HYAMP® Series

GROUND BOND TESTER

SENSE +

TEST

SENSE -

My Menu / Select

POWER

DATA

BARCODE

CONT.

CHECK

RESET

TEST

Hypot® Series

DIELECTRIC WITHSTAND TESTER

RETURN

HIGH VOLTAGE

My Menu / Select

POWER

DATA

BARCODE

CAUTION

HIGH VOLTAGE

5KVAC MAX.

6KVDC MAX.

CAUTION

HIGH VOL

TAGE

5KV

AC MAX.

6KVDC MAX.

N

L

6
7

 8

6

4

2

5

3

Summary of Contents for HYAMP 3240

Page 1: ...hould always know when a test is being performed WARNING THISGUIDEWASCREATEDFOROPERATORSHAVINGSOMEFAMILIARITYWITHELECTRICALSAFETY TESTING AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK TO PREVENT ACCIDENTAL INJURY OR DEATH THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN HANDLING AND USING A TEST INSTRUMENT CONTACT US AT INFO ARISAFE...

Page 2: ... Voltage with Interlock Disable Key unplugged Push the POWER button to turn the instrument ON Home Screen WARNING LOCATE A SUITABLE TESTING AREA WITH A THREE PRONG GROUNDED OUTLET BE SURE THAT YOUR THREE PRONG OUTLET HAS BEEN TESTED FOR PROPER WIRING READ THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST INSTRUMENT SETUP 2 To Grounded Power Source instrument Instrument will not output Fin...

Page 3: ...OWN LEFT RIGHT DRAG DROP Drag and drop meters on your Perform Test screen Simply touch hold and drag meter to any of the other meter locations Simple swipes allow you to quickly move through screens parameters TOUCH HOLD Touch and hold test file or test step to show available Quick Actions Save As Delete Insert etc ...

Page 4: ...Swipe up and down to scroll through test parameters Select Next to continue through each parameter to adjust values based on your application Press the Test Screen button to exit out of Parameters Press ENTER to save your file changes You will be redirected to the Test Screen TAP TEST SETTING Home Screen TAP SETUP TESTS File Screen TAP ADD FILE EDIT PARAMETERS Tap parameter values to edit individua...

Page 5: ...nner so you can automatically scan product barcodes and run test sequences 1 2 3 4 5 6 Connect the black return test lead P N 38490 to the black Return terminal located on the front panel of the instrument Connect the black sense lead to the Sense terminal located just above the Return terminal Clip the end of the black return test lead to the chassis ground of the DUT Connect the red high current...

Page 6: ...L USB REMOTE INTERFACE GND T10A 250V 100 120 200 240V 850VA MAX 50Hz 60Hz INPUT SENSE CURRENT RETURN SENSE OUTPUT GND T10A 250V 100 120 200 240V 360VA MAX 50Hz 60Hz REMOTE INTERFACE INPUT OUTPUT RETURN H V CONT CHECK SIGNAL OUTPUT SIGNAL INPUT CAUTION HIGH VOLTAGE SIGNAL OUTPUT SIGNAL INPUT CAL USB REMOTE INTERFACE GND T10A 250V 100 120 200 240V 850VA MAX 50Hz 60Hz INPUT SENSE CURRENT RETURN SENSE...

Page 7: ... 3 4 5 Plug the DUT line cord into the adapter box Once the test parameters are set start the test from the HYAMP If the Ground Bond test passes the Hipot test will automatically start If a test fails the test sequence will abort and failure results are displayed RETURN CURRENT RESET TEST HYAMP Series GROUND BOND TESTER SENSE TEST SENSE My Menu Select POWER DATA BARCODE CONT CHECK RESET TEST Hypot...

Page 8: ... HYAMP please consult the full Operation and Service Manual or call us toll free 1 800 858 TEST 8378 or 1 847 367 4077 2017 Associated Research arisafety com 7 2017 Safety Is Our Only Focus FOLLOW US Via Acquanera 29 22100 Como tel 031 526 566 r a fax 031 507 984 info calpower it www calpower it ...

Reviews: