3.2
Operation of the ASSKEA pro
wound
®
M and ASSKEA pro
wound
®
S
3.2.1
Operation at initial startup
1.
Press the
button for 1-2 seconds to switch on the
ASSKEA pro
wound
®
M or
ASSKEA pro
wound
®
S
. The following start screen is displayed for 5 seconds:
2.
The
Setup
menu is displayed:
3.
Use the
arrow buttons to select the
Language
menu.
4.
Use the
button to confirm your choice.
5.
Use the
arrow buttons to select the desired language:
6.
Use the
button to confirm your choice.
7.
Select the prescribed operating mode in the
Setup
menu.
8.
Use the
button to confirm your choice.
9.
Perform the desired treatment (please refer to chapter 3.2.2 or 3.2.3).
10.
Switch off the
ASSKEA pro
wound
®
M or ASSKEA pro
wound
®
S
by pressing the
button
for 3 seconds.
3.2.2
Continuous operation
In the continuous operating mode, the
ASSKEA pro
wound
®
M and ASSKEA pro
wound
®
S
maintain a continuous vacuum in the wound dressing.
The vacuum pressure value can be set from -20 mmHg to -200 mmHg in steps of 5 mmHg.
The default value is -80 mmHg.