Page 36 of 38
Handling of electromagnetic interactions
8.2
Although electromagnetic interference of the
ASSKEA secretion aspirator proVisio
®
M28
has
been reduced to a minimum, electromagnetic disturbances between the
ASSKEA secretion
aspirator proVisio
®
M28
and other devices cannot be excluded. Therefore you should comply
with the specified requirements and instructions regarding the permissible electromagnetic
environment at any case and monitor the
ASSKEA secretion aspirator proVisio
®
M28
in order
to ensure proper operation and to prevent adverse events for patients and users. Select another
location for the
ASSKEA secretion aspirator proVisio
®
M28
, if the permissible electromagnetic
environment cannot be ensured or if you have observed functioning of the
ASSKEA secretion
aspirator proVisio
®
M28
or of another device in the vicinity that is not intended.
Since electromagnetic propagation is affected by absorption and reflection from structures,
objects and people and since the field strength from fixed transmitters cannot be predicted in
advance with accuracy, a site survey of the electromagnetic phenomena at the location at which
the
ASSKEA secretion aspirator proVisio
®
M28
shall be operated should be considered to
assess the existing electromagnetic environment at this location. If unusual performance is
detected, additional measures may be required, for example a modified orientation or another
location of the
ASSKEA secretion aspirator proVisio
®
M28
. If the essential features of the
ASSKEA secretion aspirator proVisio
®
M28
are impaired by electromagnetic disturbances, a
reduced or non-existent flow rate and vacuum may be expected.
List of all cables and transformers replaceable by the
8.3
user
Name
Specification
Maximum length
Power
supply unit
incl.
connecting cable
Type: GTM91099-6015-3.0-T2
Manufacturer: GlobTek
Technical data:
100-240 V~, 50-60 Hz, 1.5 A (in)
12 V DC, 5 A (out)
1.20 m
Mains cable
Type: H03VVH2-F
Manufacturer: HAWA
Technical data: 250 V~, 2.5 A
1.80 m