Askle Sante WINNCARE automorpho AUTOMORPHO User Manual Download Page 11

 

 

- 9 - 

 
J

 

ALARMS 

 
If there is a 

pressure alarm

, the red indicator will flash and an intermittent audio signal will sound. The audio signal can be cancelled 

by pressing the 

 button. 

 

1.

 

Check the feed pipe connection to the compressor and that the CPR valve is closed 

2.

 

Check that the air lines are positioned and connected correctly and not split 

3.

 

Check that there are no leaks in the cells (structure, connection) 

4.

 

Replace defective parts if necessary (pipes, cells, etc.) 

5.

 

If the problem persists, contact your reseller 

When the pressure returns to normal, the compressor will revert to normal operation and the indicator will go out. 
 
If there is an 

electrical fault alarm

, the red indicator will flash and a continuous  audio signal will sound. The audio signal can be 

cancelled by pressing the 

 button. 

 

1.

 

Check that the compressor is still connected (to the pump and to the power point) and that the mains indicator is on. 

2.

 

If the problem persists, we advise you strongly to disconnect the air lines, blank them off with the transport plug and move the 
patient regularly. 

 
When  the  power  supply  returns,  you  must  restart  the  compressor  by  pressing  the  Start/Stop  button  while  repeating  the 
INSTALLATION/USE instructions, §.

I

, starting from step 8, after you have reconnected the mattress feed pipe to the compressor. 

 
 

 COMMENTS: 

-

 

On  starting  the  compressor,  there  is  an  initialisation  step.  If  there  is  a  fault,  an  alarm  is  indicated  by  rapid  flashing  of  all  the 
indicators or, depending on the type of problem by the servicing indicator flashing.

 

-

 

If the compressor is disconnected, even temporarily, there is always an electrical fault alarm to warn against accidental handling. 

 

 WARNING:

 

After accidental deflation of the mattress due to a prolonged power cut, check at the time of reflation that the patient does not have a 
limb wedged between the mattress and the bed sidebars. 
 
 
 
 

CPR (Emergency Cardio-Pulmonary Resuscitation)  
 

In case of cardiac arrest:

 

1.

 

Turn the end of the CPR valve to set the dial to OPEN, 

2.

 

The mattress will deflate and the back of the thorax will be in less than 15 seconds on the firm base of the bedframe to allow 
External Cardiac Massage.  

 
 

Summary of Contents for WINNCARE automorpho AUTOMORPHO

Page 1: ... PA0010085 MAJ 08 01 2013 UK User manual ...

Page 2: ......

Page 3: ......

Page 4: ...CTOGRAMS 3 C DESCRIPTION 3 D HOW IT WORKS 3 E INDICATIONS 4 F PRECAUTIONS 4 G SPECIFICATION 5 H CONTROL PANEL 6 I INSTALLATION USE 8 J ALARMS 9 K MAINTENANCE DISINFECTION 10 L MAINTENANCE SERVICING 10 M STORAGE 12 N DISPOSAL OF THE MEDICAL DEVICE 12 O WARRANTY 12 ...

Page 5: ...foam base The cells are inflated alternately one out of two except for the 3 head cells which remain static The whole is protected by an integral removable waterproof and breathing cushion incorporating a pillow D HOW IT WORKS The alternating pressures prevent prolonged vascular compression which may lead to histohypoxia The low pressure static mode enables you to handle patients who need to be im...

Page 6: ... European Directive 93 42 CEE relating to the essential criteria applicable to medical devices only compatibility between systems assembled by the manufacturer ASKLESANTE guarantees a safe combination when using the AXTAIR motorised air mattress The specification and performance of the motorised air support will only be maintained by using the AXTAIR pump ref VAXT4 POMPE AUTO associated with the A...

Page 7: ...d inflation pump Compatible ref VKIT AXT SPECIFICATION OF THE AUTOMORPHO COMPRESSOR Alternating mode 1 cell out of 2 Alternates every 7 minutes Low pressure static mode Automatic pressure control dependent upon the patient s morphology Manual optimisation in semi seated position 50 Static mode for care closed plan Flow 8 litres min Mattress inflation time Less than 20 mins for an Automorpho Approx...

Page 8: ...initial inflation Servicing indicator This orange indicator lights up continuously if servicing of the compressor is advisable The indicator flashing together with an alarm sound indicates a problem with the distributor In this case please contact your reseller See L MAINTENANCE SERVICING Mains indicator This orange indicator when on permanently indicates that the compressor is connected to the ma...

Page 9: ...ettings This zone is used to set the parameters Press the corresponding button to adjust the settings The blue lights come on continuously to validate the selection Pressure button this runs a pressure calculation with an AUTOMORPHO compressor at any time A steady light means that the calculation is OK Pressure level indicator low medium or high This button adjusts the pressure according to the po...

Page 10: ...ndicators come on alternately For a precise calculation the patient should not move excessively The calculation lasts a maximum of 6 minutes When the calculation is finished the pressure button indicator goes out and the pressure indicators are lit continuously to indicate that the pressure is now matched to the patient s morphology The compressor then starts DYNAMIC mode automatically 11 Selectin...

Page 11: ...onnect the air lines blank them off with the transport plug and move the patient regularly When the power supply returns you must restart the compressor by pressing the Start Stop button while repeating the INSTALLATION USE instructions I starting from step 8 after you have reconnected the mattress feed pipe to the compressor COMMENTS On starting the compressor there is an initialisation step If t...

Page 12: ...nence time The removable foam base can be replaced in case of alteration Ref VAXT3 MS support x4 Base 5cm CE marking compulsory under directive 93 42 CEE for products claiming that they are suitable for the surfaces of Medical Devices Cover Observe the following washing conditions Wash in water maximum temperature 90 C reduced mechanical action rinsing in decreasing temperature spin drying Bleachi...

Page 13: ...ré Traitement de surface Décontamination Répartition de surface par µfibres Solution base alcoolique EPI masque Hygiène gants tablier Propreté physique Inacceptable Entretien de surface Méthode 1 Opération Produit Détergent Désinfectant EPI gants tablier Lavage housse en machine à 90 C Détergent Housse de remplacement identique à la protection d origine Contrôle Qualité Conforme Matelas et Compres...

Page 14: ...ctronic equipment waste at the end of its working life You can contact the store reseller or distributor from whom the product was bought to find out the collection points for worn out devices closest to your home Before any disposal the appliance must be cleaned in accordance with the instructions in K SERVICING DISINFECTION to avoid any risk of contamination We are grateful for your co operation...

Page 15: ......

Page 16: ...ASKLÉSANTÉ 200 rue Charles Tellier ZI du Grézan 30 034 Nîmes Cedex 1 France Tél 33 0 4 66 02 15 15 Fax 33 0 4 66 02 15 00 Email contact askle com www askle com ...

Reviews: