4. THE TILTABLE ARM DOES NOT MOVE AT ALL
CAUSE
SOLUTION
Counterweights are situated in a
higher weight than this of the person
on the scale.
Put counterweights in a lower weight.
The transmission rod is loose.
Hook the transmission rod in the lever
extension (figure 2).
Transmission rod is connected in a
wrong way.
Hook transmission rod properly
(figure 2).
Graduated rule is in blockage position
(upwards)
Turn down the blockage of the
graduated rule.
5. LA BÁSCULA ESTÁ DESCALIBRADA
CAUSE
SOLUTION
Maladjusted during transport
Contact your distributor or Customer
Service of
asimed
.
Mortises are worn-out because of
prolonged use
Contact your distributor or Customer
Service of
asimed
NOTE
:
In order to maintain the exactness of the graduated rule and
prolong its life, manipulate the big counterweight lifting it up without
touching the mortises.