Technical Guideline 

In-Line Purge Pump Filter 

Replacement Instructions 

TG-95-0400 

TG-95-0423 

Rev B | 01/12/21 

Page 5 of 5

 

 

Tech Support Website: ASIsupport.com | Parts Order Website: ASI.parts | Tech Support:  [email protected] or 303.407.6072 

 

 

 WARNING 

PRESENCE OF HEAVY METALS/AMALGAM 

This system may be equipped with optional suction instruments. As part of dental procedures, particles of amalgam may be 

suctioned into the dental suction handpieces and collected within the system and trap filter. 

 

 

 WARNING 

INFECTIOUS MATERIALS 

Infectious disease workplace safety protocols to safeguard against cross contamination of infectious disease should always 

observed. When maintaining the suction system or emptying the contents of the suction waste container, safe precautions 

and practices including the wearing of face mask, eye protection and gloves are to be followed. 

 

Summary of Contents for TG-95-0400

Page 1: ...200933 may easily retrofitted with the In Line Purge Pump Filter by purchasing the purge pump filter PN 95 0400 The replacement screen should be replaced at a minimum quarterly 95 0400 In Line Purge Pump Filter Assembly OR 95 0423 Replacement Screen Filter Qty of 4 CONTENTS Parts Included 1 Tools Required 1 Instructions 1 Warnings Cautions 4 PARTS INCLUDED In Line Purge Pump Filter Assembly 1 OR R...

Page 2: ...cal cord out of the way of the back door Fig 1 Fig 1 3 To gain access to the purge tubing assembly first remove the steel reinforced vacuum pump tubing by unscrewing the tubing clamp and pulling the tubing from the brass fitting Fig 2 Fig 2 4 Before removing the entire tubing assembly from the unit it may be necessary to manually drain any waste from the suction canister Locate the quick connect f...

Page 3: ...ain from the suction canister completely Fig 4 If only replacing the filter screen skip Step 6 Fig 4 6 Locate the quick connect fitting where the purge tubing meets the sediment bowl Using a flat head screwdriver press on the quick connect fitting then pull the tubing away from the fitting to completely remove the tubing assembly from the unit Fig 5 Fig 5 7 If replacing the filter screen unscrew t...

Page 4: ...sons or animals in any case when the device is misused or not operated applied or maintained in strict accordance with user owner instructions set out in the operation manual In the event of any doubt or question the user is to contact ASI for clarification or assistance Improperly maintained or operated systems or instruments may void the associated warranties WARNING COMPRESSED AIR The compresse...

Page 5: ...ion instruments As part of dental procedures particles of amalgam may be suctioned into the dental suction handpieces and collected within the system and trap filter WARNING INFECTIOUS MATERIALS Infectious disease workplace safety protocols to safeguard against cross contamination of infectious disease should always observed When maintaining the suction system or emptying the contents of the sucti...