background image

– 16 –

mmH

2

O

mPa

mBar/cmH

2

O

kPa

(Gauge)

(Gauge)

(Gauge)

(Gauge)

3000

1

250

25

5000

1.6

300

40

10,000

2.5

400

60

6

500

100

10

600

160

40

1000

250

1600

400

2000

600

2500

1000

4000

5000

6000

10,000

psi

psi

psi

bar/kb/cm

2

bar

(Gauge)

(Compound)

(Absolute)

(Gauge)

(Compound)

vac.

–15/+15

15

1

–1/0

5

–15/+30

25

1.6

–1/1

10

50

2.5

–1/2

15

4

–1/3

30

6

–1/30

60

10

100

16

160

25

200

40

300

60

500

160

600

250

800

400

1000

500

1500

2000

2500

3000

5000

7000

DIGITAL  TEST  GAUGE  RANGES

I&M008 10080 Rev. E_layout  10/14/13  9:35 AM  Page 16

Summary of Contents for Digital Test Gauge

Page 1: ...Ashcroft Inc 2013 I M008 10080 10 13 Rev E GBE Digital Test Gauge Operating Instructions...

Page 2: ...2...

Page 3: ...stry of 66 high Other industry leading fea tures include twelve selectable engineering units seven languages and password protected disable and calibration functions With the range printed on the keyp...

Page 4: ...4...

Page 5: ...ions CONFIG Mode Units Engineering 9 Update Rate 9 10 Auto Off 10 Backlite 10 Languages 10 Damping 11 Contrast 11 Calibrate Gauge Calibration 11 Default Password 12 Zero Calibration 13 Zero Span 13 Ze...

Page 6: ...r maximum displayed values UNITS Pressure ENGLISH PSI INHG INH2O with Temperature Options 60 F 4 C 20 C FTSW METRIC BAR MBAR KPA MPA mmHG CMH2O mmH2O KG CM2 UPDATE RATE Pressure measurement per second...

Page 7: ...charge Flashing display when unit pressured beyond full scale Press to turn backlite on or off Range on keypad complies with ASME B40 7 Press to turn unit on or off AUTO OFF Turns unit off after optio...

Page 8: ...auge To clear minimum and maximum values press ZERO CLR button when min max values are indi cated This key allows for customization of the gauge Pressing the CONFIG key allows cycling through the main...

Page 9: ...g the rate at which pressure is updated on the display screen The default rate measures pressure at the maximum rate of 10 updates per second or 100 milli seconds Optional rates of measurement are mea...

Page 10: ...g back lite button on will decrease battery life To use the BACKLITE option Step 1 Press the CONFIG key until the word BACKLITE appears Step 2 Press the ENTER key Step 3 Press the CONFIG key to select...

Page 11: ...1 2 3 4 5 6 and 7 Step 1 Press the CONFIG key until the word CONTRAST appears Step 2 Press the ENTER key Step 3 Press the CONFIG key to select one of the available CONTRAST options Step 4 Press the E...

Page 12: ...rd Step 10 Press the ENTER key Step 11 Decide on a five number user password then follow the procedure above insert ing a number in the flashing display until all five numbers are inserted Step 12 A S...

Page 13: ...djustment of full scale range an ERROR message will be displayed when the ENTER button is pressed If this occurs press the ZERO CLEAR button on the keypad to return to the previous screen 2 ASME B40 7...

Page 14: ...back to the unit of measure of the gauge The DISAB feature disables the zero feature of the gauge Zero Disable Feature This feature allows disabling the Zero Clear button on the keypad It also allows...

Page 15: ...tep 2 Press the ENTER key Step 3 Insert the user password following the procedure as described in steps 3 through 13 above Note This is the same user password as in the CALIBRAT mode Step 4 Press the...

Page 16: ...40 1000 250 1600 400 2000 600 2500 1000 4000 5000 6000 10 000 psi psi psi bar kb cm2 bar Gauge Compound Absolute Gauge Compound vac 15 15 15 1 1 0 5 15 30 25 1 6 1 1 10 50 2 5 1 2 15 4 1 3 30 6 1 30 6...

Page 17: ...00 9 00 Ranges Vac thru 7000 psi see engineering units below for other units of measurement Operating Temperature 0 150 F Storage Temperature 40 180 F DISPLAY Type LCD Display Digits 5 Character Heigh...

Page 18: ...ns On Off 10 Sec 30 Sec 5 Min 15 Min Field Recalibration Zero Span password protected Contrast 7 Available Options Disable Config Options Allows disabling of Config Options password protected Calibrat...

Page 19: ...Installation and Replacement The gauge comes standard with three AAA alkaline batteries installed For battery replacement use only one of the three types listed below Energizer AAA alkaline E92 LR03 A...

Page 20: ...Ashcroft Inc 250 Main Street Stratford CT 06614 5145 Tel 203 378 8281 Fax 203 385 0602 e mail info ashcroft com www ashcroft com Ashcroft Inc 2013 I M008 10080 10 13 Rev E GBE...

Reviews: