Installation and Maintenance Instructions for           

28HPX High Purity Gauge

© Ashcroft Inc. 2020, 250 East Main Street, Stratford, CT 06614-5145, USA, Tel: 203-378-8281, Fax: 203-385-0357, www.ashcroft.com 

All sales subject to standard terms and conditions of sale. 28_HPX_Gauge_im_RevA_(I&M011-10276)_04-03-20

2

   

3-3 

Unpacking 

Handle product with great care during unpacking. 
Verify the product has not been damaged and that the 
instrument fulfills the required specifications. Please con-
tact the dealer or Ashcroft if an issue exists.

4.   DRAWING

   4.1  Unpacking 

Displacement occurs in the diaphragm once the pressure 
is applied via the process connection. The pressure is then 
transmitted to the movement with the pressure being indi-
cated by the movement of the pointer.

5.   INSTALLATION 

 

5.1 

Install the instrument where the environment is free from 

vibration, humidity, dust and high temperature.

  

5.2

 The pressure gauge is packaged in a hermetically-sealed 

polyethylene bag. To avoid contamination, only remove the 
unit from the bag 

immediately before

 use/installation.

  

5.3

 Take care not to damage the diaphragm surface.

  

WARNING

Scratches or scoring of the swivel coupling can result in 

leaking of the measuring device. This may prove hazard-

ous, though this would depend upon the type of applica-

tion being measured.

  

5-4

 Install the pressure gauge where it is held in a stable 

position.

  

5-5

 Utilize a wrench to tighten the process connection cou-

pling nut. Only apply a wrench to the coupling flats; no force 
should be applied to the gauge housing.

  

CAUTION

Do not apply force to the gauge housing as this can result 

in damage and adversely affect product performance.

 

5-6

 Reduce pressure pulsation to an absolute minimum.

6.   OPERATION 

6.1 

Before applying pressure to the gauge, make sure that 

the pointer is positioned at “0” pressure. 

6.2 

Raise and lower pressure slowly, while avoiding sudden 

increase/decrease of pressure.

 

6.3 

Never apply pressure in excess of the unit’s pressure 

rating.

   6.4 

Use the pressure gauge for measuring pressure less than 

½ to 

2

3

 of the full-scale range.

 

  

WARNING

Never apply pressure beyond the maximum pressure (i.e., 

the full-scale value). Injury and/or damage may occur if 

pressure element ruptures due to extreme pressure in 

excess of its pressure rating.

   6.5 

Measured fluids, gas or liquid, should be compatible and 

not corrosive to SUS316L.

  

WARNING

Do not use this pressure gauge with fluids that may  

corrode the instrument’s wetted parts.

If corrosive fluid flows through this pressure gauge, the 

pressure element (diaphragm) may be damaged or burst 

and it flows out, resulting in injury or damages to the 

peripheral device and equipment.

 

7. MAINTENANCE

   7.1 

Check the gauge accuracy regularly to ensure correct 

indication; this is recommended to be once or more per year.

   7.2

 Pressure indication error of approximately 1 graduation is 

repairable, though units must be replaced when the error is 
identified as substantial.

8.   TROUBLESHOOTING 

Refer to the accompanying table and implement appropriate 
countermeasures. Contact Ashcroft in the event the issue 
persists.

   8.1 

Refer to the accompanying table and implement 

countermeasures. Please contact Ashcroft in the event the 
issues persist.

Pointer

Connection

Socket

Diaphgram

Movement

Dial

Summary of Contents for 28HPX

Page 1: ...ents 4 Secondary damage caused by the non conformance of the products including the above Regardless of recognized mishandling any apparent evi dence of deformation abrasion burnout or other identified issue shall be excluded from the warranty scope with the cost being incurred by the customer Definition of Safety Terms Safety precautions within this manual have been classified as follows DANGER I...

Page 2: ...uge housing CAUTION Do not apply force to the gauge housing as this can result in damage and adversely affect product performance 5 6 Reduce pressure pulsation to an absolute minimum 6 OPERATION 6 1 Before applying pressure to the gauge make sure that the pointer is positioned at 0 pressure 6 2 Raise and lower pressure slowly while avoiding sudden increase decrease of pressure 6 3 Never apply pres...

Page 3: ...ears due to vibration 3 Movement wears due to pressure fluctuations 1 Re install pressure gauge in a different position 2 Re install pressure gauge in a different position 3 Reduce pressure fluctuations The pointer does not read zero when the pressure gauge is removed from the piping 1 Check whether excessive pressure had been applied 2 Check whether excessive vibration or fluctuating pressure had...

Page 4: ... 28HPX High Purity Gauge Ashcroft Inc 2020 250 East Main Street Stratford CT 06614 5145 USA Tel 203 378 8281 Fax 203 385 0357 www ashcroft com All sales subject to standard terms and conditions of sale 28_HPX_Gauge_im_RevA_ I M011 10276 _04 03 20 4 ...

Reviews: