14
TROUBLESHOOTING:
Problem:
Power is 'ON', but lamp does not
illuminate:
Light intensity appears dimmer than
normal:
Fuse is blown:
Correction:
- Turn intensity up.
- Replace fuses, 0.160 Amp, 5 x 20mm, 250V, slo-blow
- Check plug and electrical source.
- Turn intensity up.
- Clean Lens Cover and/or Fresnel Lens. (Note: Before
cleaning, turn power to Lamp ‘Off’.)
- Check plug and electrical source.
- LED Assembly might be damaged - Return entire Lamp
Assembly to Aseptico for repairs.
- Replace both fuses, 0.160 Amp, 5 x 20mm, 250V,
slo-blow