14

TROUBLESHOOTING:

Problem:

Power is 'ON', but lamp does not

illuminate:

Light intensity appears dimmer than

normal:

Fuse is blown:

Correction:

- Turn intensity up.
- Replace fuses, 0.160 Amp, 5 x 20mm, 250V, slo-blow
- Check plug and electrical source.

- Turn intensity up.
- Clean Lens Cover and/or Fresnel Lens. (Note: Before

cleaning, turn power to Lamp ‘Off’.)

- Check plug and electrical source.
- LED Assembly might be damaged - Return entire Lamp

Assembly to Aseptico for repairs.

- Replace both fuses, 0.160 Amp, 5 x 20mm, 250V, 

slo-blow

Summary of Contents for ALU-40MOM

Page 1: ...OPERATION AND MAINTENANCE INSTRUCTION MANUAL ALU 40MOM PORTABLE DENTAL LAMP ALU 40MOM PORTABLE DENTAL LAMP...

Page 2: ...487 3157 Toll Free 800 426 5913 www aseptico com info aseptico com INFORMATION CONCERNING THE ACCURACY AND PRECISION OF THIS PRODUCT MAY BE OBTAINED UPON REQUEST BY CONTACTING ASEPTICO AT THE ADDRESS...

Page 3: ...ions provided in this manual to receive the best and longest service from your Aseptico equipment Separate manuals may be provided to cover the operation and maintenance of other accessories for your...

Page 4: ...able to a properly grounded outlet only CAUTION The ALU 40MOM Lamp is sensitive to shock and may be damaged if dropped or impacted against a hard surface WARNING Do not disassemble or alter the Lamp s...

Page 5: ...e strong enough to safely support the lamp 2 Loosen the boom adjustment handle and extend boom to desired length and or pivot the lamp head to the desired operating position Tighten handle to secure t...

Page 6: ...try of usage may need to be purchased separately Optional DC Power If the power source is 12VDC or 24VDC use the optional DC Battery Clip or Adapter Assemblies see Figure 3 to connect to the DC source...

Page 7: ...ble brightness control When viewed from the rear of the Lamp Head rotate Ring clockwise to increase lamp intensity or counter clockwise to decrease intensity The intensity adjustment range is from 0 t...

Page 8: ...r shelf on the Front Housing see Figure 7 e Attach Lens to Housing with the three Fasteners from Step 1c above f Replace Lens Cover 2 LENS ADJUSTMENT Follow these steps to adjust the position of the L...

Page 9: ...sh soap to a lint free lens tissue and gently wipe Lens Clean the Lens Cover using the same method Reinstall Fresnel Lens and Lens Cover onto Lamp Head b Metal Components The metal exterior may be cle...

Page 10: ...8 ALU 40MOM PARTS DIAGRAM Figure 9 Lamp Boom ALU 40MOM LAMP HEAD Refer to Figure 11 page 12 1 2 3 4 5 6 7 9 11 10 8 12 13 5 14 11...

Page 11: ...OC M6 X 8MM BLK 510731 1 HEAT SHRINK 1 8 BLK 870118 1 HEAT SHRINK 1 4 BLK 870120 2 5 HOLE PLUG NYLON 1 2 IN BLACK 510202 1 M S STNLS PHDPHL 10 24 X 5 8 THRD CUTTING 510732 1 END CAP MODIFIED BOOM ALU...

Page 12: ...10 ALU 40MOM PARTS DIAGRAM Figure 10 Lamp Power Supply 15 16 20 23 24 42 19 18 37 21 16 33 36 28 35 34 26 27 28 23 25 32 31 29 30 40 38 39 45 43 44 41 17 22...

Page 13: ...UL RING 6 22 16 860004 1 WIRE PVC 22 GRN YEL 54 870260 0 21 CONN JACK DC POWER PCB MOUNT 860333 1 FUSE 5 X 20 SLO BLO 0 160 AMP 830046 2 FUSE DRAWER 2 POLE 840060 1 POWER INLET FUSED SWITCHED 840059 1...

Page 14: ...12 Figure 11 Lamp Head 2 4 5 8 9 18 17 6 3 6 7 10 11 13 19 16 15 1 21 12 20 14...

Page 15: ...1 WASHER SHIM 43 OD X 115 ID X 01 STNLS 510726 1 APERTURE ADJUST RING ALU 40MOM 461931 1 RES 100K TRIM POT 350V 1 2W 810430 1 WIRE 24 TEFLON 41 40 TINNED COPPER WHT 870314 2 WIRE 24 TEFLON 41 40 TINNE...

Page 16: ...up Replace fuses 0 160 Amp 5 x 20mm 250V slo blow Check plug and electrical source Turn intensity up Clean Lens Cover and or Fresnel Lens Note Before cleaning turn power to Lamp Off Check plug and el...

Page 17: ...74 cm Adjustable 27 56 in 70 cm focal length Beam Characteristics Color temperature 6 500 Kelvin Light Output 1 500 ft c 24 Typ 16 145 lux 61 cm Typ Light Distribution 7 different aperture sizes Nomi...

Page 18: ...tion consult accompanying documents Type B Equipment Do Not Throw Into Trash Class II Equipment Manufacturer Authorized European Representative Protective Earth Ground Temperature Limitation Atmospher...

Page 19: ...ter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing...

Page 20: ...P N 420905 Rev I ECO 14684 03 2019 PRINTED IN THE USA P O Box 1548 Woodinville WA 98072 8333 216th Street S E Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com...

Reviews: