Aseptico AEU-16C TORQUE PLUS+ Operation And Maintenance Instruction Manual Download Page 5

5

OPERATION FUNCTIONS:

1. Power Switch 

- Controls power on/off

to console. When turned on, console

should light up.

2. Speed Control 

- Controls the

motor/handpiece speed. Turn clock-

wise for increased speed, counter-

clockwise for decreased speed. The

L.E.D. speed display indicates relative

speed from minimum to a maximum of

35,000 RPM.

3. Mode Switch 

- Sets the unit to acti-

vate from foot switch or from power

switch. If using the foot switch, set the

mode to 'FOOT'. Set the mode to

'HAND' to activate whenever the power

switch is on.

4. Rotation Switch 

- Selects clockwise

or counter-clockwise rotation of the

bur. Set the rotation to 'FWD' for nor-

mal clockwise operation and to 'REV'

for counterclockwise operation.

5. Optional Foot Switch 

- Controls the

on/off activation of the motor/hand-

piece. Variable-speed foot switches

are also available. Plug the foot switch

into the foot switch receptacle on the

back of the console and select 'FOOT'

mode for proper operation.

Summary of Contents for AEU-16C TORQUE PLUS+

Page 1: ...ic Lab Motor System OPERATION and MAINTENANCE INSTRUCTION MANUAL P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 International 425 487 3157 Toll Free 800 426 5913 www asept...

Page 2: ...Weight 4 lbs 1 8 kg Power Source 110 120V 60Hz or 220 250V 50Hz Max Power Consumption 190 W ENVIRONMENTAL CONDITIONS Designed for indoor use only Altitude up to 2000 m Temperature range of 5 C to 40...

Page 3: ...ork Congratulations This system is engineered to provide many years of reliable service Please read the instructions provided in this manual to receive the best and longest service from your Aseptico...

Page 4: ...cord to the back of the console and plug into a grounded electrical receptacle 3 Connect the motor to the receptacle on the lower front panel of the console OPTIONAL AE 14A VARIABLE SPEED FOOT CONTRO...

Page 5: ...foot switch or from power switch If using the foot switch set the mode to FOOT Set the mode to HAND to activate whenever the power switch is on 4 Rotation Switch Selects clockwise or counter clockwis...

Page 6: ...osi tion The front panel display will light 2 Set the Mode and Rotation as desired 3 Turn the Speed Control knob until the desired operating speed is shown on the Speed Display 4 When mode is set to H...

Page 7: ...lution MOTOR HANDPIECE The included motor handpiece assembly is NOT auto clavable Do not attempt to disassemble the motor or motor connector Do not oil or lubricate the motor Do not attach a handpiece...

Page 8: ...6 Changing Motor Brushes 1 Unscrew the motor end 2 Pull the plug out of motor end 3 Unscrew the motor brushes 4 Replace Brushes and align as shown 5 Replace plug and screw on end cap...

Page 9: ...7 Serial Number Protective earth ground Speed revolutions Increase Dangerous Voltage Attention consult accompanying documents SYMBOL DEFINITIONS SN...

Page 10: ...50V slo blo fuse for 110 volts 5 A 250V slo blo fuse for 220 volts Check motor plug connection Increase RPM Check that a bur is properly seated in the handpiece and the collet is closed If foot switch...

Page 11: ...9 NOTES...

Page 12: ...discretion to repair or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to...

Reviews: