6. Drain and disconnect the intake hose. Disinfect all hoses which have come into contact with waste

water using a dilute bleach solution.

7.  Store  the  intake,  cold  water  and  drain  hoses  along  with  the  alternate  source  regulator  in  the

plumbing pouch. Be sure to coil tubing carefully to avoid kinking. Dry components as thoroughly

as possible before storage.

8.Disconnect the ON/OFF foot switch, power cord, and instrument tray. Store these items in the

electronics accessories pouch. Disconnect optional soap dispenser if used and flush out the soap

spout. Pack and store dispenser components in pouch. 

9.  Twist  loose  and  remove  the  four  sink  legs,  or  optional  leg  supports  if  used.    Fold  and  store

components in their appropriate pouches.

10. Place the sink into the case with the sink sides aligned with the case handles. Store the plumbing

and electrical pouches, and optional soap dispenser pouch if used, in the sink basin. Lower the

sink spout, or slide the Locking Cap up the tube and pivot the optional 15-inch spout forward until

it lies flat with top surface of basin.

11.Store the leg and optional leg-support pouches in the lid. Place the lid and secure all perimeter

latches. 

Reference figure A

Daily cleaning

All hoses which have come in contact with waste water during operation must be disinfected with a

dilute bleach or equivalent solution, prior to storage. Disinfect exposed sink parts in accordance with

conventional medical/dental practice.

Daily functional checklist

Observation of the following conditions will ensure that the ADU-40CF is functioning normally.

1. The green L.E.D. by the spout lights when the unit is connected to a power source.
2. The pump can be heard running when the ON/OFF footswitch is depressed.  
3. The spout is fully extended.
4. When the tank is filled the orange L.E.D. lights indicating the heating element is operating.
5. Once primed, the intake lines retain their water level when the pump is not operating.

Summary of Contents for ADU-40CF Command Air

Page 1: ...ADU 40CF Portable Field Sink NSN 6520 01 456 5380 OPERATION MAINTENANCE SERVICE MANUAL COMMANDlAIR ...

Page 2: ...ures A1 A2 Case Contents 4 Figures B1 B2 Legs Pouches 4 Figure C Electronics Pouch 4 Figure D Optional Soap Dispenser Pouch 4 Figure E Plumbing Pouch 4 Figure F Optional Leg Supports 5 Figures G1 G2 Spout Positions 5 Figure H Connections 5 Figure I Spout Connection 5 Figure J Optional Soap Dispenser 6 Figure K Regulator Assembly 6 Figures L1 L2 Spout Guides Removal 11 Figure M Ground 11 Figure N F...

Page 3: ...inted metal Sink Intake Water Connection options NATO standard jerry can Alternate source Water bladder max 10 psi Established pressurized water source Sink Waste Water Connection NATO standard jerry can Sink flow rate 0 68 gpm 2 2 lpm Max pump discharge head 8 3 ft 2 5 meters Pump self priming limit 48 in 1 2 meters Heating capacity 1500 Watts Tank capacity 0 75 Gal 2 9 Litre Peak water temp 106 ...

Page 4: ... power cord Foot switch Instrument tray Case Contents Figs A1 A2 Sink assembly 1 MIL Spec carry case 1 Operation Service Manuals 2 Accessory pouches 3 Optional Soap Dispenser pouch 1 Optional Leg Support Kit pouch 1 1 2 3 4 1 2 3 4 1 2 Figure B1 Legs Pouch Figure C Electronics Pouch 1 2 3 4 Figure A2 Open Case with Optional Spout Optional Accessories Pouches 1 2 3 4 1 2 3 4 3 4 1 2 1 2 1 2 3 5 5 6...

Page 5: ...cure the line using the attached push on cap Refer to Figure H 6 Attach the waste hose to the underside of the sink Place the opposite end in a waste water jerry can not provided Secure the line using the push on cap Refer to Figure H 7 If desired attach the components of the cold water pump assembly refer to Figure E to dispense cold water a Unpack the manual foot pump and tubing assembly b Attac...

Page 6: ... internal pump Connection to this type of source can be made using the existing intake tube For connection to an established water source with an operating pressure above 5 psi the regulator assembly is required Attachment is performed by disconnecting the intake hose and replacing it with the regulator assembly Refer to Figure K A line from the water supply having male garden hose threads can the...

Page 7: ...r cold water Water will automatically be warmed as it flows through the internal water tank The outlet temperature will vary with consumption and source temperature As a general rule a full tank of room temperature water will take two and a half to three minutes to warm to the shut off temperature of 106 deg Fahrenheit 41 deg Centigrade Optional Soap Dispenser Position the optional soap dispenser ...

Page 8: ...nal soap dispenser pouch if used in the sink basin Lower the sink spout or slide the Locking Cap up the tube and pivot the optional 15 inch spout forward until it lies flat with top surface of basin 11 Store the leg and optional leg support pouches in the lid Place the lid and secure all perimeter latches Reference figure A Daily cleaning All hoses which have come in contact with waste water durin...

Page 9: ...quire lubrication or have special cleaning instructions Performance verification and Inspection No specific inspection schedule needs to be maintained for this unit Components have been selected for their maintenance free service life The drain catch should be cleaned after each use Verifying that the following conditions are satisfied during operation will ensure proper function These conditions ...

Page 10: ...own Inspect control board and heating element replace components as necessary Heating element damaged Replace element Not heating water orange LED not lit Spout not raised Raise spout Tilt switch shut off Ensure sink is standing upright Circuit board thermal sensor Inspect control board and components replace as required Level sensor Replace level sensor Not heating water green L E D not lit No sy...

Page 11: ...s 7 Reassemble by reversing steps Once the interior of the sink has been accessed the following services can be performed Replacing fuses The control board and heating element are protected by the 15A fuse and the pump is protected by the 2A fuse Both these fuses are located on a bracket atop the solenoid valve The fuses are not intended to be operator replaceable A blown fuse is a symptom of a mo...

Page 12: ...ing its tabs in opposite directions Use pliers to improve clamp manipulation Work the tubing off each end of the pump 2 Loosen the green retaining screw and remove the grounding wire from the pump 3 Remove the two nuts which mount the pump to the bracket 4 Trace and disconnect the pump wiring from the control board and the solenoid valve Connect the new wires at the old locations 5 Replace the pum...

Page 13: ...f the spout guide 5 Replace the o ring and reassemble by reversing steps Be careful not to deform the spout tube when reinstalling the roll pin Replacing seal in optional 15 inch Spout Refer to Figure L2 Tools Needle nose pliers small screw driver or dental instrument 1 Remove spout from the tank as per Spout Guide Removal 2 Remove three capscrews from bottom of spout base with Allen wrench 3 Remo...

Page 14: ...Page 14 NOTES ...

Page 15: ...Page 15 NOTES ...

Page 16: ...P N 420203 Rev Q ECO 14703 04 2019 P O Box 1548 Woodinville WA 98072 1548 1 800 426 5913 425 487 3157 Fax 360 668 8722 email info aseptico com Internet www aseptico com ...

Reviews: