6. Drain and disconnect the intake hose. Disinfect all hoses which have come into contact with waste
water using a dilute bleach solution.
7. Store the intake, cold water and drain hoses along with the alternate source regulator in the
plumbing pouch. Be sure to coil tubing carefully to avoid kinking. Dry components as thoroughly
as possible before storage.
8.Disconnect the ON/OFF foot switch, power cord, and instrument tray. Store these items in the
electronics accessories pouch. Disconnect optional soap dispenser if used and flush out the soap
spout. Pack and store dispenser components in pouch.
9. Twist loose and remove the four sink legs, or optional leg supports if used. Fold and store
components in their appropriate pouches.
10. Place the sink into the case with the sink sides aligned with the case handles. Store the plumbing
and electrical pouches, and optional soap dispenser pouch if used, in the sink basin. Lower the
sink spout, or slide the Locking Cap up the tube and pivot the optional 15-inch spout forward until
it lies flat with top surface of basin.
11.Store the leg and optional leg-support pouches in the lid. Place the lid and secure all perimeter
latches.
Reference figure A
Daily cleaning
All hoses which have come in contact with waste water during operation must be disinfected with a
dilute bleach or equivalent solution, prior to storage. Disinfect exposed sink parts in accordance with
conventional medical/dental practice.
Daily functional checklist
Observation of the following conditions will ensure that the ADU-40CF is functioning normally.
1. The green L.E.D. by the spout lights when the unit is connected to a power source.
2. The pump can be heard running when the ON/OFF footswitch is depressed.
3. The spout is fully extended.
4. When the tank is filled the orange L.E.D. lights indicating the heating element is operating.
5. Once primed, the intake lines retain their water level when the pump is not operating.